TS1 Patient Monitor 6 Parameters

Introduction

The monitor is used to monitor, display, check, store, alarm and transmit many physiological parameters, including ECG (3/5/12 lead, arrhythmia analysis, ST segment analysis and HR ), RESP, SpO2, PR, NIBP and TEMP,IBP etc. It is applicable for use by adult, children and neonate.

The monitor can be used in medical institutions or during transport monitoring inside/outside the hospital environment, it is used by clinical professionals or under their guidance or professionals via training, it is not suitable for use in family.


Function

1)5" touch screen, easy and convenient to operate, automatic lock screen function prevents misoperation.

2)Portable design (weight < 1.0 kg), flexible to transport.

3)Multi - interface selection, convenient to observe patient’s data trend.

4)Degree of protection provided by enclosure complies with IEC/EN 60529, waterproof degree: IP44

5)Fanless design, quiet, energy-saving and clean, which reduces the possibility of cross-infection.

6)Built-in large-capacity battery, meets the requirements of long-term operation.


Performance

(1)ECG 

Lead type:

3-lead: I, II, III 

5-lead: I, II, III, aVR, aVL, aVF, V

12-lead: I, II, III, aVR, aVL, aVF, V1,V2,V3,V4,V5,V6

Waveform

3-lead: 1 channel

5-lead: 2-channel 

10-lead: 12-channel

Gain: 2.5 mm/mV, 5.0 mm/mV, 10 mm/mV, 20 mm/mV

Scan speed: 12.5 mm/s, 25 mm/s, 50 mm/s

HR

Measurement and alarm range:

Adult: 15 ~ 300 bpm

Pediatric/neonate: 15 ~ 350 bpm

Resolution: 1 bpm

Accuracy: ± 1 % or ± 1 bpm, whichever is greater

Alarm error: ± 2 bpm

ST-segment detection

Measurement and alarm range: - 2.0 mV ~ + 2.0 mV

Accuracy: - 0.8 mV ~ + 0.8 mV: ± 0.04 mV or ± 10 %, whichever is greater

Other range: unspecified

Arrhythmia analysis: 13 kinds

Pacemaker: yes

(2)RESP

Measurement mode: R-F (RA-LL) impedance

Respiration rate: 

Measurement range: 0 ~ 150 rpm

Resolution: 1 rpm

Accuracy: 7 ~ 150 rpm, ± 2 rpm or ± 2 %, whichever is greater; 

 0 ~ 6 rpm, unspecified

Apnea alarm: 10 ~ 40 s

Scan speed: 12.5 mm/s, 25 mm/s

(3)NIBP 

Measurement mode: Oscillometry

Working mode: Manual/AUTO/Continuous 

Measurement interval in AUTO mode: 1 / 2 / 3/ 4 / 5 / 10 / 15 / 30 / 60 / 90 / 120 / 240 /480 /960 minutes

Measurement period in Continuous mode: 5 minutes

Measurement and alarm range: 10 ~ 270 mmHg

Resolution: 1 mmHg

Cuff pressure accuracy: ± 3 mmHg

Measurement accuracy:

Maximal mean deviation: ± 5 mmHg

Maximal standard deviation: 8 mmHg

Over-pressure protection:

Adult mode: 297 ± 3 mmHg

Pediatric mode: 240 ± 3 mmHg

Neonatal mode: 147 ± 3 mmHg

(4)SpO2

Measurement and alarm range: 0 ~ 100%

Resolution: 1 %

Measurement accuracy: 70 % ~ 100 %: ± 2 %; 

0 % ~ 69 %: unspecified 

(5)PR

Measurement and alarm range: 25 ~ 250 bpm

Accuracy: ± 2 bpm or ± 2 %, whichever is greater

(6)TEMP

Channel: dual-channel

Measurement and alarm range: 0 ~ 50 ℃

Resolution: 0.1 ℃

Accuracy: ± 0.1 ℃

(7) IBP

Number of channels: dual channel

Label: ART PA CVP RAP LAP ICP P1 P2

Measurement and alarm range: -50 ~ 300 mmHg

Resolution: 1 mmHg

Accuracy: ±2 % or 1 mmHg, whichever is greater

(8) Power supply: 100V-240V~ 50Hz/60Hz (Adapter output DC 12V)

(9) Safety type: Class I CF type anti-defibrillation application part

(10) Waterproof grade: IP44


Accessories 

Category Accessory name 

Main unit TS1 main unit 

ECG cable 5-lead, American standard, TPU

SpO2 probe (integrated) Digital fingertip SpO2 probe for adult  (CMS-N-SPO2 6P, 3M, yellow)

TEMP probe R25=2.252K TEMP probe, skin surface type, CMS-N-TEMP 2P, PVC, L=3M

NIBP cuff BP extension tube, L=3M

Adult cuff (25 ~ 35 cm)

IBP module


Physical characteristic

Dimension: 142 mm × 67 mm × 102 mm

Weight: < 1 kg



Buy safe Product

The following FDA Disclaimer is required for all eBay listing in Healthcare category and is included for REFERENCE: The sale of this item may be  subject to regulation by the U.S. Food and Drug Administration and state and local regulatory agencies. If the item is subject to FDA  regulation, we will verify your status as an authorized purchaser of  this item before shipping of the item.


The Fingertip Pulse Oximeter is registered on the Australian Register of  Therapeutic Goods (ARTG)with the code 197923, and certified by FDA of  United States and CE,TUV of Europe. The  Fingertip Pulse Oximeter that  is FDA 510K Approved

Please use under the doctor's instruction

Shipping:

1). We will ship within 1 business day after your payement ;ship from China warehouse by China Post 2-4 weeks Arrival

2).We don't add taxes, VAT or other hidden charges. You pay us what you see  on the invoice, i.e. goods subtotal + shipping cost (not include  duties). Please find out as much as you can about your import taxes in  your own country before bidding the item. You need to pay for the import duties in case that occurred for certain goods, or you may kindly give  us some suggestions about that, we promise to do our best to help you!

About Warranty:

We guarantee new equipment other than accessories to be free from defects  in workmanship and materials for a period of twelve months (six months  for spare parts) from the date of shipment under normal use and service. Our company's obligation under this warranty is limited to repairing,  at our company's option, any part which upon our company's examination  proves defective.

Return Policy:

If you receive a defective item which you want to return, please contact  us within 5 days from the day you receive the item. All return items  must be returned with its original packaging and accessories. If we  confirm the damage happened during the shipment, we will bear the  shipping cost for return.

We will refund the money to you when we get the return items. Or replace item for you free of charge. 

ibay365