Capnograph CO2 Patient Monitor ETCO2 Vital Signs 7 Parameters,FDA/CE

               
Introduction
The monitor has abundant  functions that can be used for clinical monitoring with adult, pediatric and  neonate. Users may select different parameter configuration according to  different requirements. The monitor, power supplied by 100-240V~,50/60Hz,   adopts 12.1'' color TFT LCD displaying real-time date and waveform. It can  synchronously display eight-channel waveform and full monitoring parameters  equipped with an optional 48mm thermal recorder. The monitor can be connected to  the central monitoring system via wire or wireless network to form a network  monitoring system.
This device can monitor such parameters as ECG, RESP,  NIBP, SpO2, and dual-channel TEMP, etc. It integrates parameter measurement  module, display and recorder in one device to form a compact and portable  equipment. Its replaceable internal battery brings lot of convenient for patient  moving.

Functions
Standard  parameters: ECG, RESP, SpO2, PR, NIBP, dual-channel TEMP,
ETCO2
1)ECG   Heart  rate(HR)
      ECG waveform
      Arrhythmia and ST-segment  analysis
2)RESP   Respiration rate(RR)
      Respiration  waveform
3)SpO2   Oxygen saturation(SpO2)
      Plethysmogram(PLETH)  waveform
      Pulse rate(PR)
      Bar graph
4)NIBP   Systolic  pressure(SYS), Diastolic pressure(DIA), Mean pressure(MEAN)
5)TEMP   T1, T2,  TD
6)IBP(optional)  CH1:SYS,DIA,MAP
CH2:SYS,DIA,MAP
IBP  waveform
7)CO2  EtCO2
InsCO2:Inspired Minimum CO2
AwRR:Air  Way Respiration
It has abundant functions, such as audible and visual alarm,  trend data storage and output, NIBP measurement, alarm event marking and drug  concentration calculation, etc.

Features
1)12.1'' TFT color LCD,  multi-language interface(Simplified Chinese, Traditional Chinese, English,  French, German, Turkish, Spanish, Portuguese, Italian, Dutch, Romanian, Russian,  Kazakh, Polish, Czech).
2)Fanless design, quiet, energy-saving and clean,  which reduces the possibility of cross-infection.
3)All-round monitor for  adult, pediatric and neonate.
4)With standard interface, oxygen graph, trend  graph, big character interface and view bed, convenient to observe.
5)Finish  all operations by keys and knobs.
6)Maximum 8-channel waveform synchronous  display.
7)Display 7-lead ECG waveform on one screen, cascade ECG waveform  display
8)Adopt digital SpO2 technology, anti-motion and anti-ambient light  interference, and measurement can be performed under the circumstance of weak  filling.
9)Heart rate variability (HRV) analysis function
10)NIBP  measurement mode: Manual/AUTO/STAT, storage for 4800-group NIBP  data.
11)Review for 71 alarm events of all parameters and 60 arrhythmia alarm  events.
12)Drug concentration calculation and titration table  functions.
13)One-touch printing of trend graph
14)Connect to Central Monitoring System by 3G, Wi-Fi or wired mode.
15)AC/DC, built-in rechargeable  lithium battery achieve uninterrupted monitoring.
16)Anti-high frequency  surgical unit, defibrillation-proof (special leads are  necessary).

Performance
▲ECG
Lead mode: 3-lead or  5-lead
Lead selection: I, II, III, aVR, aVL, aVF, V
Waveform: 5-lead, dual-channel
3-lead, single-channel
Gain: 2.5mm/mV, 5.0mm/mV, 10mm/mV,  20mm/mV, 40mm/mV
Scan speed: 12.5mm/s, 25 mm/s, 50 mm/s
HR: 
Measurement and alarm range: 15~350 bpm
Accuracy: ±1 % or ±1 bpm, whichever is greater
Alarm accuracy: ± 2 bpm
Resolution: 1  bpm
ST-segment monitoring:
Measurement and alarm range: -2.0 mV ~ +2.0  mV
Accuracy: -0.8 mV~+0.8 mV  ±0.04 mV or ±10%, whichever is greater 
Other range: unspecified
Arrhythmia analysis: ASYSTOLE, VFIB/VTAC,  COUPLET, BIGEMINY, TRIGEMINY, R ON T, VT>2,
PVC, TACHY, BRADY, MISSED  BEATS, PNP, PNC
Pacemaker: yes
▲RESP
Method: R-F(RA-LL) Impedance 
Respiration rate:
Measurement and alarm range: 0~150 rpm
Resolution:  1 rpm
Measurement accuracy: ±2 rpm
Alarm accuracy: ±3 rpm
Apnea alarm:  10~40s
Scan speed: 6.25 mm/s, 12.5 mm/s, 25 mm/s
▲NIBP
Method:  Oscillometry
Mode: Manual/AUTO/STAT
Measurement interval in AUTO mode:  1/2/3/4/5/10/15/30/60/90/120/240/480/960 minutes
Measurement period in STAT  mode: 5 minutes
Measurement and alarm range: 10 ~ 270 mmHg
Resolution: 1  mmHg
Cuff pressure accuracy: ±3 mmHg
Measurement accuracy:
Maximal mean  deviation: ±5 mmHg
Maximal standard deviation: 8 mmHg
Over-pressure  protection:
Adult mode: 297±3 mmHg
 Pediatric mode: 240±3  mmHg
 Neonatal mode: 147±3 mmHg
▲SpO2
Measurement and alarm range: 0 ~  100%
Resolution: 1%
Measurement accuracy: 70%~100%: ±2%;
0%~69%:  unspecified
▲PR
Measurement and alarm range: 30 ~ 250 bpm
Measurement  accuracy: ±2 bpm or ±2%, whichever is greater
▲TEMP
Channel:  dual-channel
Measurement and alarm range: 0 ~ 50℃
Resolution: 0.1  ℃
Accuracy: ±0.1 ℃
▲EtCO2
Method: Sidestream or Mainstream 
Measuring Range: 0~150mmHg
Resolution:
0~69 mmHg, 0.1 mmHg
70~150  mmHg, 0.25 mmHg
Accuracy:
0~40 mm Hg ±2 mm Hg
41~70 mm Hg ±5%
71~100  mm Hg ±8%
101~150 mm Hg ±10%
AwRR Range: 2~150 rpm
AwRR Accuracy:  ±1BPM
Apnea Alarm: Yes
▲IBP
(optional)
Channel: dual-channel
Label: ART, PA,  CVP, RAP, LAP, ICP, P1, P2
Measuring and Alarm Range: -10~300  mmHg
Resolution 1 mmHg
Accuracy: ±2% or 1mmHg, whichever is  greater
▲Power supply: 100-240V~, 50/60Hz
▲Safety classification: Class I,  type CF defibrillation-proof applied part

Accessories
1)Adult fingertip SpO2 probe  (5-pin)
2)Adult NIBP cuff
3)NIBP extension tube
4)ECG lead  cable
5)ECG electrode
6)Temperature probe
7)Power cord
8)User  Manual
9)ETCO2

Physical characteristic
Dimension: 314 mm(L) × 145 mm(W) × 264  mm(H)
Weight: 3.9 Kg

Buy Safe Product!
The following FDA Disclaimer is required for all eBay listing in Healthcare category and is included for REFERENCE: 
The sale of this item may be subject to regulation by the U.S. Food and Drug Administration and state and local regulatory agencies. If the item is subject to FDA regulation, We will verify your status as an authorized purchaser of this item before shipping of the item.

Shipment

1. We will ship the item within 48 hours after receiving your payment, please kindly leave your detail shipping address and telephone number, both are very important for the shipment,thank you.

2.We will upload the tracking number after shipment,you can search from both ebay and paypal.You can track online.
3. If you haven't received the item after 2 weeks of your payment, please contact me immediately! 

Terms of Sale

1.If you have a defective item, you want to return or discount. Please contact us within 60 days from you receive the shipment.All return items must be returned with it's original packaging and accessories. Customer is responsible for shipping charges on returned items.
2.we will refund the money to you when we get the return items. Or replace item for you. We will deal with the refund in 2 business days when we get the info from you.Please read our trade Policy before buy , if you don't agree our trade policy please don't buy it. 
               
           
       

       
   

Only English user manual,if you need any other language,please contact us!ibay365