Introduction
The device adopts sidestream method to measure the animal’s parameter related to EtCO2 and AwRR. It can detect pulmonary ventilation function and reflect the circulation and pulmonary blood flow and indirectly reflect the alveolar ventilation.
It is applicable for use in pre-hospital emergency, emergency department, ICU, operating room, respiratory department and other occasions.
Features
1)End-tidal CO2 measurement function;
2)Respiration rate measurement function;
3)Optional three units;
4)With audible and visual alarm and silencing functions, the alarm limit can be set;
5)Over-limit alarm function for EtCO2, Respiration rate;
6)Low battery alarm function;
7)With the prompt function of “Checking Filterline”;
8)User can calibrate zero by themselves.
9)With atmospheric pressure and temperature compensation functions
10)Rechargeable lithium battery, up to 8-hour user time;
Performance
End-tidal CO2 measurement:
1)Measurement range: 0 ~150mmHg
2)Resolution: 1mmHg
3)Accuracy:
0~40 mmHg:±2 mmHg
41~70 mmHg:±5% of readings
71~100 mmHg:±8% of readings
101~150 mmHg:±10% of readings
Respiration rate measurement:
1)Measurement range: 2rpm~120rpm
2)Resolution: 1rpm
3)Accuracy: ±1rpm
Safety characteristic:
1)Device type: internally powered equipment
2)Degree of protection against electric shock: type BF applied part
3)Degree of protection against ingress of liquid: IPX1
Other parameters:
1)Preheat time: 2min;
2)Response time: < 3s;
3)Alarm sound pressure level: 50 dB;
4)Battery life time: 8hours;
5)Gas sampling rate: 60 ml/min ~ 100 ml/min.
Accessories
User Manual
Sampling tube(Cannula tube with filter cotton)
USB cable
Physical Characteristic
Dimension: 159*72*31 mm
Weight: 270 g
Working environment
Temperature: 5℃~40℃
Relative humidity: 30%~ 75%, no condensation
Atmospheric pressure: 700hPa~1060hPa
Power supply: DC 3.7V (rechargeable lithium batteries)
Input power: ≤1.5VA