• PC Interface

  • The serial USB (Universal Serial Bus) port is used for communication between a PC and the HI-PRO USB unit.

    Communication:

    SB 2.0 full Speed ( USB 1.1 compatible)

    USB Connector Type

    “ Type B” connector (oil the HI-PRO USB unit)

  • Power Supply

    Rated Voltage:

    4.50v-5.25v

    Maximum Current Consumption( active)

    < 500 mA(2.5 W)

    Power Consumption during USB Suspend

    < 500 uA (2.5 mW)

     

  • Output ratings

  • The following output ratings are valid for Left and Right hearing instrument connectors.
  • Fixed Battery supply (pin 1), common for Left and Right side

    1.35V, 10/50mA(current rating is controlled by fitting software)

    Programmable battery supply (pin 5), common for Left and Right side

    -3.50V to +3.50V 30 mA(voltage is controlled by fitting software)

  • Operating environment
  • Operation at temeratures below -20°C or above +60°C may cause permanent damage.
  • Temperature

    5°C-40°C

    Humidity

    30 to 90%

    Warm-up time

    <20 seconds

    Air pressure

    600 hPa to 1060 hPa

  • Dimension and Weight
  • Size (L*W*H)

    137*114*37mm

    Net Weight

    230g

  • Patient Safety

  •  The HI-PRO USB unit complies with the following standards:

     EN 60601-1, Type BF

    UL60601 and CAN/CSA-C22.2 NO 601.1-90

    To comply with the above standards, the programming cable and the connector to the hearing instrument must meet the following requirements:

     No conductive parts may be accessible when the programming cable is connected too HI-PRO USB unit.

     The programming cable and the connector must provide double isolation and be able to withstand a dielectric strength test potential of 500V.

  • Standards
  • Safety:

    EN 60601-1, Type BF

    Electromagnetic compatibility:

    EN 60601-1-2