i1 Pro 2 Rev E Spectrometer EFI-ES2000

I just turned it on to run xrite diagnostics on it.

i1Pro Tests

General Information
  Date and Time: 08/03/2024 09:54:35
  Application version: 4.2.5.381, 32-bit
  SDK version i1Pro: 4.2.4.5061
  SDK version i1iO: 2.2.4.5061
  SDK version i1Pro3: 1.0.15.5919
  SDK version i1iO3: 1.0.15.5919
  SDK version i1iSis: 2.0.0.4968
  SDK version ColorMunki: 1.0.3.4630
  SDK version i1Display2: X-Rite, Inc. SipCal Library Ver. 7.3.7 (32 bit)
  Platform: Microsoft Windows 10, 64-bit

Open instrument
  Connection: Successful
  Open instrument: Successful

Instrument information
  Instrument inspection: Successful
  Serial number i1Pro: 7043778
  Type: i1Pro
  Revision: E
  Firmware version: 634
  CPLD version: 999
  Dark measurement count: 127754
  Lamp burning time: 66.4 seconds
  Spot measurement count: 291
  Scan measurement count: 0
  Factory calibration: 06/08/2015
  Spectral Module ID: 01-96d19817000060
  Supplier: 'EFI-ES2000'
  Recertification date: -

Instrument Licenses
  Measurement Modes: [Reflectance Spot, Reflectance Scan, Emission, DualReflectance Spot, DualReflectance Scan]
  Licensed functionality: [Has UV LED, Has Wavelength LED, Has Zebra Ruler sensor, Max Ruler Length = 300 mm, Has Indicator LED, Spectral values, Tristimulus, Density]
  Licensed Modules: 
  i1Match Module "Scanner": NOT licensed
  i1Match Module "Monitor": NOT licensed
  i1Match Module "Printer": NOT licensed
  i1Match Module "Printer RGB only": NOT licensed
  i1Match Module "Printer CMYK only": NOT licensed
  i1Match Module "Beamer": NOT licensed
  i1Match Module "Printer Easy RGB": NOT licensed
  i1Match Module "Printer Easy CMYK": NOT licensed
  i1Match Module "Digicam": NOT licensed
  i1Match Module "Editor": NOT licensed
  i1Profiler Module "Display Profiling": NOT licensed
  i1Profiler Module "RGB Printer Profiling": NOT licensed
  i1Profiler Module "CMYK Printer Profiling": NOT licensed
  i1Profiler Module "Printer QA": NOT licensed
  i1Profiler Module "Advanced-Mode": NOT licensed
  i1Profiler Module "Projector": NOT licensed

Button test
  Button: Successful

Emission test
  Emission calibration: Successful
  Prepare the instrument for emission measurement
  White measurement: Successful
  Gray measurement: Successful
  Black measurement: Successful
  Red measurement: Successful
  Green measurement: Successful
  Blue measurement: Successful
  Emission: Successful

Button test
  Button: Successful

Noise Tests
  Noise measurements: Successful
  Noise result: Successful

Reflectance test
  Reflectance Spot calibration: Successful
  Reflectance Spot measurement 1: Successful
  Reflectance Spot measurement 2: Successful
  Reflectance Spot measurement 3: Successful
  Reflectance Spot measurement 4: Successful
  Reflectance Spot measurement 5: Successful
  Reflectance Spot measurement 6: Successful
  Reflectance Spot measurement 7: Successful
  Reflectance Spot measurement 8: Successful
  Reflectance Spot measurement 9: Successful
  Reflectance Spot measurement 10: Successful
  Lamp test: Successful
  Reflectance drift test: Successful
  Reflectance Scan calibration: Successful
  Dual Reflectance Spot calibration: Successful
  Dual Reflectance Spot test: Successful
  Dual Reflectance Scan calibration: Successful

Positioning Sensor Test
  Scan with positioning sensors: Successful (Moving distance: 25.3 cm)

*** PASS ***


**************************

i1Pro Tests

General Information
  Date and Time: 2024-03-08T09:54:35
  Application version: 4.2.5.381, 32-bit
  SDK version i1Pro: 4.2.4.5061
  SDK version i1iO: 2.2.4.5061
  SDK version i1Pro3: 1.0.15.5919
  SDK version i1iO3: 1.0.15.5919
  SDK version i1iSis: 2.0.0.4968
  SDK version ColorMunki: 1.0.3.4630
  SDK version i1Display2: X-Rite, Inc. SipCal Library Ver. 7.3.7 (32 bit)
  Platform: Microsoft Windows 10 (Build 19041), 64-bit

Open instrument
  Connection: Successful
  Open instrument '\\?\usb#vid_0971&pid_2000#6&1c4d2f9b&0&17#{989d6515-e122-4442-a357-751457f102a5}': Successful

Instrument information
  Instrument inspection: Successful
  Serial number i1Pro: 7043778
  Type: i1Pro
  Revision: E
  Firmware version: 634
  CPLD version: 999
  Dark measurement count: 127754
  Lamp burning time: 66.4 seconds
  Spot measurement count: 291
  Scan measurement count: 0
  Factory calibration: 2015-08-06
  Spectral Module ID: 01-96d19817000060
  Supplier: 'EFI-ES2000'
  Recertification date: -

Instrument Licences
  Measurement Modes: [ReflectanceSpot, ReflectanceScan, EmissionSpot, DualReflectanceSpot, DualReflectanceScan]
  Licensed functionality: [Has UV LED, Has Wavelength LED, Has Zebra Ruler sensor, Max Ruler Length = 300 mm, Has Indicator LED, Spectral values, Tristimulus, Density]
  Licensed Modules:
  i1Match Module "Scanner": NOT licensed
  i1Match Module "Monitor": NOT licensed
  i1Match Module "Printer": NOT licensed
  i1Match Module "Printer RGB only": NOT licensed
  i1Match Module "Printer CMYK only": NOT licensed
  i1Match Module "Beamer": NOT licensed
  i1Match Module "Printer Easy RGB": NOT licensed
  i1Match Module "Printer Easy CMYK": NOT licensed
  i1Match Module "Digicam": NOT licensed
  i1Match Module "Editor": NOT licensed
  i1Profiler Module "Display Profiling": NOT licensed
  i1Profiler Module "RGB Printer Profiling": NOT licensed
  i1Profiler Module "CMYK Printer Profiling": NOT licensed
  i1Profiler Module "Printer QA": NOT licensed
  i1Profiler Module "Advanced-Mode": NOT licensed
  i1Profiler Module "Projector": NOT licensed

Button Test
  Button: Successful

Emission test
  Emission calibration: Successful
  Prepare the instrument for emission measurement
  White measurement: Successful (X=97.870 Y=102.234 Z=112.203)
  Gray measurement: Successful (X=32.776 Y=32.719 Z=53.758)
  Black measurement: Successful (X=0.085 Y=0.080 Z=0.150)
  Red measurement: Successful (X=47.620 Y=25.120 Z=0.795)
  Green measurement: Successful (X=30.686 Y=66.392 Z=12.438)
  Blue measurement: Successful (X=17.974 Y=8.987 Z=98.997)
  Emission: Successful

Button Test
  Button: Successful

Noise Tests
  Noise measurements: Successful
  Noise result: Successful (electronic noise: 2.969)

Reflectance test
  Reflectance Spot calibration: Successful (slope=0.0679931045 (0.150000006), WLS=0.600000024)
  Reflectance Spot measurement 1: Successful (95.866, -0.566, 2.004)
  Reflectance Spot measurement 2: Successful (95.853, -0.580, 2.037)
  Reflectance Spot measurement 3: Successful (95.860, -0.588, 2.036)
  Reflectance Spot measurement 4: Successful (95.869, -0.615, 2.055)
  Reflectance Spot measurement 5: Successful (95.875, -0.616, 2.044)
  Reflectance Spot measurement 6: Successful (95.878, -0.623, 2.047)
  Reflectance Spot measurement 7: Successful (95.883, -0.627, 2.062)
  Reflectance Spot measurement 8: Successful (95.889, -0.629, 2.057)
  Reflectance Spot measurement 9: Successful (95.894, -0.629, 2.048)
  Reflectance Spot measurement 10: Successful (95.901, -0.643, 2.058)
  Lamp test: Successful (minL=95.853 meanL=95.877 MIT=0.0136439996 CIT=0.0136439996 LOT=70.0214005)
  Reflectance drift test: Successful (0.028)
  Reflectance Scan calibration: Successful (slope=0.0679931045 (0.150000006), WLS=0.650000036)
  Dual Reflectance Spot calibration: Successful (slope=0.0679931045 (0.150000006), WLS=0.650000036)
  Dual Reflectance Spot test: Successful
  Dual Reflectance Scan calibration: Successful (WLS=0.650000036)

Positioning Sensor Test
  Scan with positioning sensors: Successful (min: -0.000, max: 0.253, total distance: 0.253)

*** PASS ***