Introduction
RAYL is a portable X-ray machine. Operation voltage is DC24V.
This equipment is an X-ray generating equipment powered by a rechargeable lithium battery that mainly used for dental examination (teeth and mandible). It consists of high pressure generator, X-ray tube and control system. The control system includes equipment controller, power controller, user interface, beam limiting device, backscatter shielding and optional remote exposure switch.
This equipment is used to irradiate the intraoral X-rays and imaging for diagnostic use.
Technical Parameters
Power Adapter
Power Input: 100-240AC 50HZ 1.6A
Power Output: 25.2VDC 1.7A
High Pressure Generator
Output Power: 120W
Tube Voltage: 70KV
Tube current: 2.0mA
Working frequency: 70KHZ
Focus-skin Distance: 110mm or 200mm
Exposure Time: 0.01-2.0S
X-ray Tube
Maximum Working Voltage: 70KV
Maximum Working Current 24mA (CP)/12mA (half wave self-rectifying circuit)
Maximum Filament Current: 2.9A
Filament Voltage (under maximum filament current 2.9A): 2.4-3.0V
Nominal Focal Spot Value: 0.4mm
Target Surface Angle: 12°
Inherent Filtration: ≥0.75mmAI
Battery
Battery Capacity: 950mAh/21.09Wh
Battery Voltage: 22.2V
Machine Parameters
Working Mode: Intermittent loading
Package Size: 490*280*410mm
Net Weight: 1.6KG
Pacakge Listing:
1.Main Body
2.Machine Base
3.Power Charger
4.User Manual
5.Remote Exposure Switch
General Safety Guidelines
*Exposure mode: continuous exposure, intermittent exposure 30 times, waiting for the next exposure.
*The design and manufacture of this equipment is to ensure the maximum safe operating, and the operation and maintenance of the equipment must be carried out in strict accordance with the safety specifications and operating instructions in this manual.
*Only qualified personnel are allowed to operate this equipment.
*Observe the working environment that a fire extinguisher should always be installed near the equipment.
*Except battery replacement, this equipment must be installed, maintained and provided with preventive or periodic maintenance services by professionals.
*Make sure the power switch is off when the equipment is disabled.
*Disconnect the power before cleaning the equipment.
*Do not place the equipment and its components in a humid environment or liquid.
*Avoid placing this equipment near chemical and gas storage facilities.
*Do not make exposure when the equipment is charging.
Daily Use and Maintenance
*The equipment should be far away from the strong electromagnetic radiation environment.
*Regularly check whether the wire of the power charger is damaged.
*The equipment should be far away from strong acid and strong alkaline environment, and avoid contact with corrosive liquids.
*When the equipment is not in use, it should be properly stored to prevent children or non-professionals from touching the equipment.
*When the equipment displays 1 grid of power, please charge it immediately to prevent battery damage.
*When the equipment is not used for a long time, it should be ensured that the equipment is fully charged with 4 grids of power once and discharged within 3 months, and then stored after being fully charged with 4 grids of power.
*When charging, when the equipment displays 4 grids of power and does not flash, it means that the equipment is fully charged. It is recommended to remove the power charger immediately to prevent battery damage.
*It is forbidden to use the equipment to exposure in the charged state to prevent the battery from exploding and other hazards.
*Keep the surface of the equipment clean, wipe the equipment once a week to prevent liquid from entering the inside of the equipment.
*The equipment should be placed in a dry environment to prevent damage to electrical components.
*The equipment use and protection requirements must meet the relevant requirements of the local government.
*The disposal of equipment scrapped or battery scrapped must comply with the relevant requirements of the local government.