Output signal: Measuring range1 – 9.8 V
Output signal: Minimum load10 kΩ
Bakeout temperature≤ 80 °C
Operating temperature10 – 80 °C
Pressure max.3000 hPa
Stability of sensitivity (year)≤ 0.5 %
Full scale1000 hPa
Precision (Full Scale)2 %
Weight120 g
Linearity and hysteresis (Full Scale)≤ 0.5 %
MaterialStainless Steel
Measuring range1E-1 – 1.1E3 hPa
Measurement cable length50 m
Nominal diameterDN 16 ISO-KF
Zero point stability (Full Scale | Year)0.5 %
Protection degreeIP65
Temperature: Storage-40 – 80 °C
Thermal sensitivity drift≤ 0.5 %
Thermal zero point drift (Full Scale)≤ 0.5 %
Supply: Voltage V DC13 – 30 V DC
Supply: consumption max.≤ 0.2 W
Volume2 cm³