Temperature calibrator analog thermocouple calibration instrument K, N type



HU520 is a small milli signal generator that can generate two kinds of voltage signals (-10mV~+110mV and -100mV~+1100mV) to calibrate thermocouples.

The module increases the output of two types of thermocouple signals, as long as the input temperature is converted to the corresponding voltage output.

The module is simple to use, and the data is input by key operation, and the output is updated immediately.


Product application:

Signal generator;

Instrument calibration;



Two. Technical parameter table:

A description of the output range of the output range

Voltage 100mV -10mV~+110mV 0.01mV maximum output current +/-5mA

Voltage 1000mV -100mV~+1100mV 0.1mV

Thermocouple K -200~+1370 0.1'C

Thermocouple N -200~+1300 0.1'C



Working voltage 7~12 VDC

Working current <50 mA

Size 57*45*20 mm long X wide X height


Three, interface description and related operations

P1 interface (power supply)

VDD: power source (7 ~ 12V)

GND: power supply negative pole


P3 interface (output signal)

CH1: channel 1, producing -10mV~+110mV.

CH2: channel 2, producing -100mV~+1100mV.

GND: power supply negative.





Four. Operation Description:

(1) key Description:

UP: on the page;

DOWN: next page;

ADD: self - added key;

SUB: self reducing key;

FUN: function key; (select the operating bit for setting the data)


4.1. Generate -10mV~+110mV voltage

Interface display font:

The first line "1. P1:-10mV~+110mV"

The second line "*0.01 0.00mV"

Channel 1, producing -10mV~+110mV



4.2. Generate -100mV~+1100mV voltage

Interface display font:

The first line "2. P2:-100mV~+1100mV"

The second line "*0.1 0.0mV"

Channel 2, producing -100mV~+1100mV


4.3. Generate K type thermocouple signal

Interface display font:

The first line "3.P1:K-TC"

The second line "*0.1 0.0'C"

Channel 1, K type thermocouple signal -200 ~ +1370 degree


4.4. Generate N type thermocouple signal

Interface display font:

The first line "4.P1:N-TC"

The second line "*0.1 0.0'C"

Channel 1, N type thermocouple signal -200 ~ +1300 degree