1PC AAY80-390 4OX-2 for CITY oxygen sensor Oxygen probe 4OXV CiTiceL
 
One, oxygen sensor 4OXV parameters:
Working principle: electrochemistry
Measuring range: 1-25%
Maximum overload: 30%
Output signal: 0.10±0.02mA in the air
Response time (T90): <15S
Zero current (bias): <0.6%
Linear: It can be considered linear in many cases

Recommended load resistance: 100 ohms

Shell material: ABS
Weight: <16 g
Azimuth sensitivity: <0.2%

Operating temperature range: -20℃ to +50℃
Recommended storage temperature: 0℃-20℃
Thermal transient (temperature plunge +22°C to -20°C): <23.5%
Working pressure range: ±20% of atmosphere
Pressure coefficient: <0.02%/Mbar
Pressure dynamics (60 cm water column step change): <200% signal change
Continuous relative humidity range (at 0℃ to 20℃): 5~95%RH non-condensing
Short-term relative humidity range (at 0℃ to 20℃): 0~99%RH non-condensing
Long-term output drift: <2%/month, the entire working life is usually <5%
Expected working life: 24 months in the air
 
Precautions:
1. It is forbidden to use an electric soldering iron to solder the lead wire of the sensor pin at high temperature. Please use it with the supplied pinhole!

2. It is forbidden to use glue potting with strong heat, volatilization and corrosion properties!
 
2. Application range of oxygen sensor 4OXV:
Oxygen sensors are widely used in coal mines, steel, petrochemical, medical, etc.
 
3. Typical application of oxygen sensor 4OXV:
Oxygen alarm, atmosphere analyzer
 
 
1. Packing: 1PC
2. Warranty period: 120 day