Product Application
The horizontal autoclave,which adopted the way of gravity exchange to releasing the cold air from the chamber more completely to ensure the reliable sterilization. The control system automatically adjust the steam inlet and outlet according to chamber temperature during sterilizing. This unit is ideal equipment for clinics, scientific research institutes and other organizations to sterilize surgical instruments, fabric glassed , culture media etc.
Main Features
1.Sterilizing course automatic sterilization controlled,easy to operate.
2.With drying function,suitable dressing drying.
3.With over-temperature,over-pressure auto-protect device.
4.The door opening mechanism cannot be operated until the pressure in the chamber is reduced to 0.027MPa.It can't be start on if the door doesn't close well.
5.The safety valve will be open automatically when the inner pressure over 0.24MPa,and the steam be exhausting to the water tank.
6.The power will be cut-off automatically,cut-off the water and alarm when the machine lacking water.
7.The chamber of the sterilizer is made of stainless steel.
Technical Data
Model | WS-150YDA | WS-200YDA | WS-280YDA | WS-400YDA | WS-500YDA |
Sterilizing chamber volume | 150L φ440*1000(mm) | 200L φ515*1000(mm) | 280L φ600*1000(mm) | 400L φ700*1100(mm) | 500L φ700*1300(mm) |
Working pressure | 0.22Mpa | ||||
Working temperature | 134℃ | ||||
Adjustment of temperature | 40℃-134℃ | ||||
Time for sterilization | 0-60min | ||||
Time for drying | 0-60min | ||||
Heat average | ≤± 2℃ | ||||
Power | 9KW/ 240V/380V three-phase | 12KW/ 240V/380V three-phase | 18KW/ 240V/380V three-phase | ||
Dimension(mm) | 1400×600×1300 | 1400×670×1650 | 1400×770×1780 | 1430×880×1830 | 1800×900×1820 |
Transport dimension(mm) | 1550×750×1850 | 1560×820×1850 | 1680×920×2100 | 1600×1050×2100 | 1850×1050×2100 |
G.W / N.W | 320/240 KG | 350/260 KG | 465/365 KG | 530/420 KG | 580/470 KG |