MS200 NIBP Simulator

               

Introductions

MS200 NIBP Simulator

Introduction

It is a multi - purpose test device for testing NIBP Monitor adopted Oscillometry. Concise and elegant in appearance, easy and convenient to operate, and it can generate the pressure up to 400mmHg (53.3 kPa) by the built-in air pump. With the functions of dynamic BP simulation under several conditions, static pressure calibration, automatic leak test and over-pressure relief test, which can quickly and accurately analyze each performance of NIBP Monitor.


Functions

1)Pressure leak test

2)Over-pressure relief test

3)BP simulation under standard condition

4)BP simulation under patient condition

5)BP simulation under arrhythmia condition

6)BP simulation under respiratory interference condition

7)BP simulation of neonate

8)Wrist BP simulation

9)User-defined BP simulation

10)System setting function.

11)The parameters set can be stored.

12)ON/OFF function

13)Dynamic BP simulation


Performance

BP simulation range:

Systolic pressure: 2.7 kPa (20 mmHg) ~ 40.0 kPa (300 mmHg)

Diastolic pressure: 1.3kPa (10 mmHg) ~ 33.3 kPa (250mmHg)

Pulse volume: 0.1 cc ~ 2.4 cc  

HR range: 30 bpm ~ 250 bpm; pulse volume > 2.00 cc, maximum HR: 200 bpm

HR accuracy: ± 1 bpm (except patient condition, arrhythmia and respiratory disturbance)  

Under patient condition:

Weak pulse, tachycardia, obese, geriatric:   ±(preset value 1%+1)bpm

Moderate exercise:                     ±(preset value 1.5%+1)bpm

Strenuous exercise:                    ±(preset value 3%+1)bpm

Pressure measurement:

Range: 0kPa (0 mmHg) ~ 53.3 kPa (400 mmHg)

Resolution: 0.1 kPa, 1 mmHg, 1 cmH2O, 1 inH2O, 0.1 psi

Accuracy:

0 kPa (0 mmHg) ~ 40.0 kPa (300 mmHg), ± (0.5 % of reading + 0.1) kPa or ± (0.5% of reading +1) mmHg

40.1 kPa (301 mmHg) ~ 53.3 kPa (400mmHg), ± 2% of reading

Pressure leak test

Generate a pressure up to 400 mmHg by the built-in air pump. It always displays the peak pressure and the real-time pressure. Leak rate is displayed after 10 seconds.

Over-pressure relief test

Generate a pressure up to 400 mmHg by the built-in air pump, it displays peak pressure.

Pressure source:

Output range: 6.7 kPa (50 mmHg) ~ 53.3 kPa(400 mmHg)

Resolution: 0.1 kPa, 1 mmHg, 1 cmH2O, 1 inH2O, 0.1 psi

Deviation between target pressure and actual pressure: ± 1.3 kPa (± 10 mmHg) in range of 13.3 kPa (100 mmHg) ~ 53.3 kPa (400 mmHg) with a minimum volume of 300 cc

Airtightness

Total airtightness: < 0.3 kPa/min (2.25 mmHg/min) under “OFF” state

Airtightness of built-in cuff: 0.27 kPa/min (2 mmHg/min)

BP simulation preset value configuration table


BP simulation BP preset value BP HR Pulse volume

Standard BP simulation #1 120/80(93) 80bpm 0.68cc

#2 150/100(116) 0.65cc

#3 200/150(166) 0.60cc

#4 255/195(215) 0.55cc

#5 60/30(40) 0.75cc

#6 80/50(60) 0.71cc

#7 100/65(76) 0.69cc

BP simulation under patient condition

Healthy heart 120/80(93) 75bpm 0.68cc

Weak Pulse 110/80(90) 95bpm 0.50cc

Mild Exercise 140/90(106) 120bpm 1.00cc

Strenuous Exer 140/90(106) 162bpm 1.40cc

Obese 120/80(93) 90bpm 0.50cc

Geriatric 150/110(123) 95bpm 0.40cc

Tachycardia 120/105(110) 130bpm 0.40cc

Bradycardia 120/60(80) 45bpm 1.10cc

BP simulation under arrhythmia condition

Premature Atrial Contraction1 138/53(81) 80bpm 0.70cc


Premature Atrial Contraction2 144/64(90) 83bpm

Premature Ventricular Contraction 118/61(80) 83bpm

Atrial Fibrillation and PVC 139/72(94) 91bpm


BP simulation under respiratory interference condition

spontaneous breathing1 138/65(89) 104bpm 0.60cc

spontaneous breathing2 149/65(93) 105bpm

spontaneous breathing3 112/47(68) 86bpm

controlled breathing 132/44(73) 98bpm

BP simulation of neonate

#1 35/15(22) 120bpm 0.11cc

#2 60/30(40) 0.10cc

#3 80/50(60)

#4 100/70(80)

BP simulation of wrist cuff

#1 120/80(93) 80bpm 0.50cc

#2 160/100(120)

#3 80/55(63)

User-defined BP simulation[1] - - - -

Note[1]:Support 9 user-defined BP simulation modes. The following parameters can be defined by users:2.3.1.The value of diastolic pressure should be lower than systolic pressure, pulse volume and HR can not be set to the highest values (e.g., pulse volume ≤ 2.00cc, HR: 250 bpm; pulse volume > 2.00cc, HR: 200bpm).


Accessories

One User Manual

Two fuses

One power cord


Physical characteristic

Display: 4.3" 65 K color LCD display

Power supply: AC100 ~ 240 V   50/60 Hz, <60 VA

Environment temperature: + 5 ℃ ~ + 40 ℃

Relative humidity: < 90 %, non-condensing

Altitude: ≤ 3000 m

Atmospheric pressure: 80 kPa ~ 106 kPa

Dimension: 207 (L) × 220 (W) × 121 (H) mm

Weight: about 2.1 kg

Shipment

1. We will ship the item within 48 hours after receiving your payment,  please kindly leave your detail shipping address and telephone number, both are very important for the shipment,thank you.

2.We will upload the tracking number after shipment,you can search from both ebay and paypal.You can track online.
3. If you haven't received the item after 2 weeks of your payment, please contact me immediately! 


Terms of Sale

1.If you have a defective item, you want to return or discount. Please  contact us within 60 days from you receive the shipment.All return items must be returned with it's original packaging and accessories. Customer is responsible for shipping charges on returned items.
2.we will refund the money to you when we get the return items. Or replace  item for you. We will deal with the refund in 2 business days when we  get the info from you.Please read our trade Policy before buy , if you don't agree our trade policy please don't buy it. 


Buy safe products

The following FDA Disclaimer is required for all eBay listing in Healthcare category and is included for REFERENCE: The sale of this item may be subject to regulation by the U.S. Food and Drug Administration and state and local regulatory agencies. If the item is subject to FDA regulation, we will verify your status as an authorized purchaser of this item before shipping of the item.

The Fingertip Pulse Oximeter is registered on the Australian Register of Therapeutic Goods (ARTG)with the code 197923, and certified by FDA of United States and CE,TUV of Europe. The  Fingertip Pulse Oximeter that is FDA 510K Approved

International Buyer--Please note:

Import duties,Taxes are not included in the item price.These charges are buyers' responsibility


We are the manufacturer of medical equipment for 17 years, the following are our main products:

ECG /  ECG holter   EEG / EEG Holter      B-ultrasound

Patient Monitor      Fetal doppler             Fetal Monitor 

Spo2 Monitor          Stethoscope              Medical Image....

Agent in the world!

We are look for partners in the world now!

Do you want  to become our agent in you conutry?

Join US!

Sky-pe:xuebox3

contec.med.mary # hotmail dot com


               
           
       
       
   

Only English User Manual,Please contact us if you need other language version.thank youibay365