O2 Sensor Type D-02 suitable for Draeger Oxy-Gauge single cathode, Ohmeda Oxygen Analyser.
The O2 - Scuba Diving Sensor Type D-02 has following Specifications:Galvanic gas sensor with an extra water repellent head space. All characteristics are based on conditions at 25 °C, 50% RH and 1013 hPa.
Electrical Connector | gold-plated slip rings |
Measurement Range | 0 - 100 Vol. % |
Sensor Lifetime | < 3 years @ ambient air |
Expected Operating Life | ~ 500,000 Vol.% h |
Initial Output Signal | 14.0 to 18.5 μV @ dry ambient air |
Response Time t90 | < 12 sec |
Drift | < 1 % volume O2 / month @ air, averaged across 12 months |
Linearity Error | < 3 % @ 100 % O2, applied for 5 min |
Zero Offset Voltage | < 200 μV in 100 % N2, applied for 5 min |
Operating Temperature | 0 to 45 °C |
Pressure Range | 600 to 1750 hPa |
Influence of Humidity | -0.03 % rel. O2 reading per % RH |
Interferences | none to He, H2 and CO |
Weight | approximately 20 g |
Material in Contact with Media | PA, PPS, PTFE, stainless steel |
Storage conditions
Temperature Range | recommended: 5 to 30 °C, maximum: -20 to 50 °C |
Ambient Pressure | 600 to 1750 hPa |
Humidity | up to 100 % Rh |
Shelf Life | < 6 months recommended |