Brief Introduction
The MS100 SpO2 simulator is a kind of separate SpO2 simulator, small and lightweight. It can perform a series of tests for the oximeter with simulation means and detects the correctness of the oximeter.
Since the manufacturer of oximeters may use other R-curves, some R-curves used abroad have been embedded at present.

Properties
◆ Disconnected connection between simulator probe and host, operation and test are more convenient.
◆ 262K color and 320*240 TFT, adjustable brightness level.
◆ Press the Film button to make it easier to operate.
◆ Power supply with rechargeable lithium battery, can display the charging information of the battery.
◆ Button volume can open or close.

Main
◆ Simulation of oxygen saturation;
◆ Pulse rate simulation;
◆ Preset patient condition simulation;
◆ Can test the response time of the device;
◆ Can simulate the SpO2 and pulse rate under different amplitudes;
◆ Can test the performance of the device under various sources of interference;
◆ Can select a different R-curve.
 
Main parameters
◆ SpO2
Measuring Range: 35%~100%
Resolution: 1%
Error: If SpO2 is between 75% and 100%, the error between ± 2% and ± P is the greater (P denotes the accuracy of the oximeter being tested). if the SpO2 value is between 74% and 50%, the error is greater between ± 2% and ± P; No, if the SpO2 value is below 50%.
◆ Pulse rate
Measuring Range: 25bpm ~ 250bpm
Resolution: 5bpm
Error: 1% ± 1bpm
◆ Amplitude
Measuring Range: 0.000%~20.000%
Resolution: 1% for 1,000%~20,000%; 0.1% for 0.100% to 0.900%; 0.025% for 0.000% to 0.075%.
◆ Patient condition
24 presets and 8 presets after launch. The number of conditions can be adjusted by setting the level of movement.
◆ Voltage
 DC 3.6V~4.2V
◆ Power supply
 Voltage 3.7 rechargeable lithium battery × 1
◆ Battery life
 Charge and discharge at least 500 times.

Accomplices
◆ An MS100 User's Guide
◆ A 5.0V adapter