BRONKHORST DIGITAL PRESSURE CONTROLLER F-201CV-050-ABD-11-V


NEW IS SEALED FACTORY PACKAGING


Gas Mass Flow Controllers for low flow rates


Bronkhorst® model F-201CV Mass Flow Controllers (MFCs) are suited for accurate measurement and control of flow ranges between 0,16…8 mln/min and 0,5…25 ln/min at operating pressures between vacuum and 64 bar. The MFC consists of a thermal mass flow sensor, a precise control valve and a microprocessor based pc-board with signal and fieldbus conversion. As a function of a setpoint value, the flow controller swiftly adjusts the desired flow rate.


EL-FLOW® Select series are equipped with a digital pc-board, offering high accuracy, excellent temperature stability and fast response. The main digital pc-board contains all of the general functions needed for measurement and control. In addition to the standard RS232 output the instruments also offer analog I/O. As an option, an on-board interface can be mounted to provide CANopen®, DeviceNet™, EtherCAT®, PROFIBUS DP, PROFINET, Modbus RTU, ASCII or TCP/IP, EtherNet/IP, POWERLINK or FLOW-BUS protocols. The EL-FLOW® Select design features optional Multi Gas / Multi Range functionality, providing (OEM-) customers with optimal flexibility and process efficiency.

Technical specifications


Measurement / control system

Flow range (intermediate ranges available) min. 0,16…8 mln/min

max. 0,5…25 ln/min

(based on N2)

Accuracy (incl. linearity) (based on actual calibration) ± 0,5 % RD plus ±0,1%FS

Repeatability < 0,2 % RD

Turndown ratio up to 1:187,5 (1:50 in analog mode)

Multi fluid capability storage of max. 8 calibration curves;

optional Multi Gas / Multi Range functionality up to 10 bar abs

Settling time (in control, typical) standard: 1…2 seconds

option: down to 500 msec

Control stability < ± 0,1 % FS (typical for 1 ln/min N2)

Operating temperature -10 … +70 °C

Temperature sensitivity zero: < 0,05% FS/°C; span: < 0,05% Rd/°C

Pressure sensitivity < 0,1% Rd/bar typical N2; 0,01% Rd/bar typical H2

Max. Kv-value 6,6 x 10-2

Leak integrity, outboard tested < 2 x 10-9 mbar l/s He

Attitude sensitivity max. error at 90° off horizontal 0,2% at 1 bar, typical N2

Warm-up time 30 min. for optimum accuracy

2 min. for accuracy ± 2% FS

Mechanical parts

Material (wetted parts) Stainless steel 316L or comparable

Pressure rating (PN) 64 bar abs

Process connections compression type or face seal (VCR/VCO) couplings

Seals standard: Viton®;

options: EPDM, Kalrez® (FFKM), FDA and USP Class VI approved compounds

Weight 0,6 kg

Ingress protection IP40

Electrical properties

Power supply +15 … 24 Vdc

Max. power consumption

Supply

15 V

24 V at voltage I/O

290 mA

200 mA at current I/O

320 mA

215 mA extra for fieldbus

<75 mA

<50 mA

Analog output 0…5 (10) Vdc or 0 (4)…20 mA (sourcing output)

Digital communication standard: RS232;

options: PROFIBUS DP, CANopen®, DeviceNet™, PROFINET, EtherCAT®, Modbus RTU, ASCII or TCP/IP, EtherNet/IP, POWERLINK, FLOW-BUS

Electrical connection

Analog/RS232 9-pin D-connector (male);

PROFIBUS DP bus: 9-pin D-connector (female);

power: 9-pin D-connector (male);

CANopen® / DeviceNet™ 5-pin M12-connector (male);

FLOW-BUS/Modbus-RTU/ASCII RJ45 modular jack

Modbus TCP / EtherNet/IP / POWERLINK 2 x RJ45 modular jack (in/out);

EtherCAT®/ PROFINET 2 x RJ45 modular jack (in/out)

Technical specifications subject to change without notice.






Limitation of Liability. Buyer/Bidder hereby agrees that under all possible circumstances, our  (seller) maximum liability in any and all auctions and sales transactions is hereby limited to the purchase price actually paid by the Buyer for the applicable sale or auction transaction. IN NO EVENT WILL THE SELLER BE LIABLE FOR ANY DAMAGES INCLUDING, WITHOUT LIMITATION, ANY DIRECT, INDIRECT, INCIDENTAL, SPECIAL, PUNITIVE, EXEMPLARY OR CONSEQUENTIAL DAMAGES, LOSSES RELATED TO BUSINESS INTERRUPTION, LOSS OF BUSINESS INFORMATION OR LOST PROFITS ARISING OUT OF OR IN CONNECTION WITH THE SALE OF THE GOODS OR THE AUCTION, OR OUT OF ANY BREACH OF WARRANTY, EVEN IF YOU HAVE BEEN ADVISED OF THE POSSIBILITY OF SUCH DAMAGES.