CONTEC E3

               
                   
Introduction

E3 is a kind of electrocardiograph, which is able to sample 12 leads ECG signals simultaneously and print out the ECG waveform with thermal printing system. Its functions are as follows: recording and displaying ECG waveform in auto/manual mode; measuring ECG waveform parameters automatically, and automatic analysis and diagnosis; pacing ECG detection; prompt for electrode-off and out of paper; optional interface languages(Chinese/English, etc.); built-in lithium battery, powered either by AC or DC; arbitrarily select the rhythm lead to conveniently observe abnormal heart rhythm; case database management, etc.



Function

1)High resolution thermal-array output system(8 dots/mm), no adjustment required. Frequency Response is up to 150Hz.

2)Sync collection for 12-lead ECG, support routine 12-lead and Cabrera-lead waveform display. Get high-quality ECG waveform via AC filter, DFT filter, EMG filter and Low-pass Filter of ECG signal.

3)Display 3/6/12 - lead ECG waveform, print mode, sensitivity, paper speed, filter status, etc. on one screen, convenient for contrastively diagnosing.

4)Support for auto 1×12, auto 1×12+1, auto 2×6, auto 2×6+1, auto 3×4, auto 3×4+1, auto 4×3, auto 3-2+1, auto 3-2, rhythm 4, rhythm 3, rhythm 2 and manual. The printed content contains time, paper speed, sensitivity, calibration signal, name of lead, filter state and patient’s information. Information including printed waveform length, output measurement parameter, diagnostic conclusion, superposition QRS waveform, histogram, trend chart and interval list, can be set, and with time print function, which meets different requirements.

5)With the functions of auto-measurement and auto-interpretation for routine ECG parameters, provide measurement results for HR, PR interval, P Duration, QRS Duration, T Duration, QT/QTc interval, P/QRS/T Axis, R(V5), S(V1) and R(V5)+S(V1), etc. and auto-diagnosis conclusion, which reduces the doctor’s burden.

6)Rhythm lead can be selected freely, convenient to observe abnormal heart rhythm.

7)The subject’s clinical information (such as, ID, Name, Age, Sex, Height, Weight, etc.) can be entered.

8)1000 cases can be stored by the built-in memory, convenient for case review and statistic.

9)Multi-language (Chinese, English, Turkish, Portuguese, German, Russian and Kazakh, etc.) interface and report.

10)Power supply: AC/DC (which can adapt AC 50/60Hz), with built-in rechargeable lithium battery, charging circuit, perfect over-current and over-voltage protection circuit.

11)External USB keyboard, code scanner, printer and USB storage device can be connected.



Performance

Input way: Floating and defibrillation protection

Lead: Standard 12 leads

Patient leakage current: <10µA

Input impedance: ≥100MΩ(1Hz)

Frequency response: 0.01Hz~300 Hz(-3dB)

Time constant: ≥3.2s

CMRR: >123dB

Filter: AC Filter(AC50/60 Hz), EMG Filter (25 Hz/35 Hz (-3 dB)), DFT Filter, Low-pass Filter

Recording way: Thermal printing system

Specification of recording paper: 80 mm(W)×20 m(L) high-speed thermal paper

Time base selection(paper speed):

5 mm/s, 6.25 mm/s, 10 mm/s, 12.5 mm/s, 25 mm/s, 50 mm/s, error: ±5% within 0.2s~2.0s

Gain control(sensitivity):2.5, 5,10, 20, 40, 10/5, 20/10mm/mV, accuracy is ±5%; Standard sensitivity: 10 mm/mV±0.2 mm/mV

Auto record: record setup according to auto record format and mode, automatically change leads, automatically measure and analyze.

Rhythm record: record setup according to rhythm record format and mode, automatically measure and analyze.

Manual record: record according to manual record format.

Measurement parameters: HR, PR interval, P Duration, QRS Duration, T Duration, QT/QTc interval, P/QRS/T Axis, R(V5) amplitude, S(V1) amplitude, R(V5)+S(V1) amplitude

Product safety type: Class I type CF defibrillation-proof applied part

Polarization resistance voltage: ±950 mV

Noise level: ≤12 µVp-p

ECG signal input sampling frequency: 32 kHz

Waveform data processing sampling frequency: 1 kHz

Sampling precision: 24-bit

The minimum detection signal: 10 Hz, 20 µV(peak-peak value) deflected sinusoidal signal can be detected

Pacing detection channel: all leads

Accuracy of input signal: ±5%.

Amplitude quantization: ≤5µV/LSB

Interchannel time deviation: <100 µs

Power supply:

AC: 100V~240V(50/60Hz)

DC: 7.4V/5000mAh rechargeable lithium battery

Degree of protection against ingress of liquid: IPX0

Working mode: continuous working



Accessories

1)A lead cable

2)A limb electrode

3)A chest electrode

4)A thermal recording paper

5)A power cord

6)An earth wire

7)A user manual



Physical characteristic

Dimension: 315mm(L)×200mm(W)×77mm(H)

Weight: 1.6Kg

Buy Safe Product 
The Fingertip Pulse Oximeter is registered on the Australian Register of Therapeutic Goods (ARTG)with the code 197923, and certified by FDA of United States and CE,TUV of Europe. The Fingertip Pulse Oximeter that is FDA 510K Approved

The following FDA Disclaimer is required for all eBay listing in Healthcare category and is included for REFERENCE:

The sale of this item may be subject to regulation by the U.S. Food and Drug Administration and state and local regulatory agencies. If the item is subject to FDA regulation, we will verify your status as an authorized purchaser of this item before shipping of the item.
                   
               
                                 
Picture 1 of 12
Picture 2 of 12

ibay365