Item Description
Easy to Use: Simply immerse the pH meter electrode in the solution to be tested, readings are quickly display on a large lcd display
Wide Application: Ideal instrument for any aquarium, the fishing industry, swimming pools, spa, school laboratories, food & beverage, drinking water etc
Auto Calibrated: Features in-built calibration with 2 buffer powders(included).Just press a button to finish the calibration. This pH meter is factory calibrated. Users can use it directly
Fast and Accurate Results: No need to work with those annoying color charts and test strips.Resolution: 0.01 pH, Accuracy: ±0.01 pH
Compact Portable Design: Pocket fitting pH meter that's also lightweight and compact. Total concealed fit in your bag or pocket for easy, casual carrying
Display: LCD Digital
Accuracy: 0.01
Resolution: 0.01
Operating Temperature: 0~60°C
Measuring Range: 0.00~14.00
Probe Type: Precision glass bulb electrode
Remove protective cap.
First rinse the electrode with distilled water and suck it with filter paper.
Turn the meter on by pressing “ON/OFF” key.
Immerse the pH meter electrode in the solution to be tested (should not be over the immersion line).
Stir gently and wait around 30 seconds till the reading stabilizes.
After finishing, clear the electrode with pure water, turn the meter off by pressing “ON/OFF” key.
Always replace the protective cap after use.
Turn on PH meter
Disolve each buffer solution in 250ML of distilled water
Immerse the electrode into the PH 6.86 solution,(under the temperature of 25)
Press the “CAL”(calibration)button for 5 seconds and release
Display will start flashing 6.86
Wait until the display stops flashing.
Rinse the electrode with distilled water and dry it with filter paper.
Immerse the electrode in PH 4.00 solution,
Press“CAL”button for 5 seconds,then press and release immediately second time
Reading will start flashing 4.00
Wait until display stops flashing
Rinse and dry the electrode with distilled water as before.
Place the meter back to the buffer solution to test calibration.If incorrect,please repeat the calibration
If approximate PH of your test solution is known to be above 0,calibrate the meter using 6.86 and 9.18 solution. To do this,please follow the calibration chart,The third step is the calibration method for 9.18 point.
If the tester was used or lay aside long time: recalibrate it.
If you open battery case or replace battery: recalibrate it.
The glass bulb of the electrode is fragile. Don't touch it by hard tools and fingers, suck the water on it gently when cleaning.
Normalerweise wird der Artikel in 15-20 Tagen eintreffen. Aber manchmal dauert es ein wenig länger, geht bis 20-30 Tage oder mehr.
1. Kundenzufriedenheit ist unser oberstes Ziel. Wir glauben, dass unsere Produkte so hervorragend sind. Alle Produkte sind Qualität geprüft. Sie sind neu und in einem guten Zustand, wenn an unsere Kunden ausgeliefert. Wir sind überzeugt, Sie werden mit Ihrem Kauf glücklich sein.
2. Wenn das Produkt fehlerhaft oder Schäden bei der Ankunft ist, oder falsche Produkt geliefert, kontaktieren Sie uns bitte sofort. Rückkehr angenommen innerhalb von 30 Tagen nach Lieferdatum und Artikel müssen im ursprünglichen neuen Zustand sein, nicht getragen oder in irgendeiner Weise mit angebrachten Umbauten & Wrap verändert. Ansonsten Deal ist endgültig. Rückholverschiffen muss vom Kunden bezahlt werden.
3. Bitte kontaktieren Sie uns zuerst, wenn Sie irgendwelche Probleme haben / Fragen / Interessen. Wir werden gerne alle Fragen zu lösen, die Sie in einer herzlichen und freundlichen Art und Weise haben kann.
4. Wir schätzen Ihr positives Rückgespräch , und das gleiche im Gegenzug tun. lassen Sie NICHT negatives Feedback ohne erste Kommunikation mit uns. Bitte maximal 2 Werktage für uns zur Antwort zu.
Wenn Sie mit unserem Produkt und Service zufrieden lassen Sie uns positives Rückmeldung
Wenn ein Problem auftritt, kontaktieren Sie uns sofort mit jedem E-Mail-Anfrage. Kontaktieren Sie uns einfach die "Frage an den Verkäufer eine Frage" Link auf eBay.