DC voltage range 60.000 mV to 1000.0 V, 6 ranges, basic accuracy: ±0.025 % rdg. ±2 dgt.

AC voltage range* 60.000 mV to 1000.0 V, 6 ranges, frequency response: 20 Hz to 100 kHz

Basic accuracy 45 - 65 Hz: ±0.2 % rdg. ±25 dgt. (RMS rectification, crest factor less than 3)

DC + AC voltage range* 6.0000 V to 1000.0 V, 4 ranges, frequency response: 20 Hz to 100 kHz


Basic accuracy 45 - 65 Hz : ±0.3 % rdg. ±30 dgt. (RMS rectification, Crest factor less than 3)

Resistance range 60.000 Ω to 600.0 MΩ, 8 ranges, [600.00 nS, only DT4282

Basic accuracy: ±0.03 % rdg. ±2 dgt.

DC current range 600.00 μA to 10.000 A, 6 ranges

Basic accuracy: ±0.05 % rdg. ±5 dgt.

AC current range* 600.00 μA to 10.000 A, 6 ranges

Basic accuracy 45 - 65 Hz: ±0.6 % rdg. ±5 dgt. (RMS rectification, Crest Factor 3 or less)

AC current range* (AC clamp measurement) ...


PEAK measurement DC V measurement: signal width > 4 msec (single shot), > 1 msec (repetitive)

AC V, DC/AC A measurement, etc.: signal width 1 msec or more (single shot), 250 μsec or more (repetition)

Capacitance range 1.000 nF to 100.0 mF, 9 ranges, basic accuracy: ±1.0 % rdg. ±5 dgt.

Continuity check Conduction threshold: 20/50/100/500 Ω, Response time: 10 ms min.

Diode test Open circuit voltage: 4.5 V max., Measuring current: 1.2 mA max., Forward threshold: 0.15 V to 3 V, 7 types

Frequency range AC V, DC+AC V, AC A measurement, with pulse widths of 1 μs or more (duty ratio 50%)

99.999 Hz (0.5 Hz min.) to 500.00 kHz, 5 ranges, ±0.005 % rdg. ±3 dgt.

Decibel conversion measurement Reference impedance setting (dBm), 4 Ω to 1200 Ω, 20 types

Display of dB conversion value of AC voltage to reference value (dBV)

Temperature measurement (thermocouple) K: -40.0 to 800.0 °C

Accuracy: ±0.5 % rdg. ±3 °C + temperature probe accuracy addition


Other functions Filter function (harmonic noise reduction, limited to AC600 V/1000 V range), display value hold, auto hold, maximum/minimum value display, instantaneous value (PEAK) display, relative value display, sampling setting, internal memory of measured values (400 data), auto power save, USB communication (optional) USB communication (optional), shutter to prevent incorrect insertion, decibel conversion, 4-20 mA percent conversion

Display 5-digit LCD, max. 60000 dgt.

Display update rate 5 times/s (capacitance: 0.05 to 2 times/s depending on measured value, temperature: 1 time/s)

Power supply AA alkaline batteries (LR6) × 4, continuous use time: 100 h

Dimensions and weight 93W × 197H × 53D mm, 650 g (including test lead holder and batteries)


Accessories Test lead L9207-10 × 1, Instruction manual × 1, AA alkaline batteries (LR6) × 4



Shipping

We ship using FedEx, DHL and Japan Post.



Payment

We usually ship within 5〜10 business days of receiving cleared payment.



* International Buyers - Please Note:

Import duties, taxes and charges are not included in the item price or shipping charges.These charges are the buyer’s responsibility.Please check with your country’s customs office to determine what these additional costs will be prior to bidding/buying. These charges are normally collected by the delivering freight (shipping) company or when you pick the item up do not confuse them for additional shipping charges. I do not mark merchandise values below value or mark items as "gifts" - US and International government regulations prohibit such behavior.