ABRIDGED DATA
Deuterium-filled, fast recovery, tetrode thyratron, specifically designed for operation under medical linac conditions. A reservoir operating from the cathode heater supply is incorporated.
Peak forward anode voltage ............................. 33 kV max Peak anode current ...................................... 1000 A max Average anode current:
continuous operation.....................................1.25 Amax intermittent operation .................................... 2.2 A max
GENERAL
Electrical
Cathode (connected internally
to mid-point of heater)..................................oxide coated
Heater voltage ............................................... 6.3 + 0.2 V − 0.3
Heater current ................................................... 22
Tube heating time (minimum) ............................. 5.0 min
Mechanical
Overall length........................317.5 mm (12.500 inches) max Overall diameter .....................84.12 mm (3.312 inches) max Net weight................................... 0.7 kg (1.5 pounds) approx Mounting position (see note 1)......................................... any Base ...... pin spacing as B5F; metal shell with micalex insert Top cap (see note 2)........................................... BS448-CT3
Cooling ...................................................................... natural