HANDPIECE  FOR  INDIVIDUAL  PATIENT
Specification
Working Air Pressure 180~300kPa, 250kPa preferred
Maximum Rotate Speed >250×103r/min
Grip Power 22N~45N
Noise not exceed 80dB(A)
Diameter Circular Runout not exceed 0.03mm
Acrotorque not exceed 0.05N∙cm

Usage
1. Inset the standard bur into the hole of cartridge
Push the bur ot the bottom with bur wrench and check whether the bur is uploaded properly. Before use, test the handpiece without bur for 1 to 5 seconds ,and make sure the bur working properly.
2. Suitable Bur  
Do not use a bent,damaged or non-concentric bur.Use a bur of ISO-standard:Φ1. 595-Φ1.600mm.
3. Remove and Upload of Bur  

Remove the bur : Aim at the inner hole of head cap with bur wrench, then push it vertically, as shown in the picture.
Installation of bur A: the working position diameter of the bur should be > 1 .5 mm. First insert the handle of the bur into the inner hole of shaft, as shown in the picture, the handle of the bur should be top in place.    
Installation of bur A: the working position diameter of the bur should be ≤ 1 .5 mm. First insert the handle of the bur into the inner hole of shaft, as shown in the picture, the handle of the bur should be top in place.

Product Applicability
For high speed handpiece for driling and other stomatological operation.
Product Structure Compoent
This product is pneumatic dental handpiece. The main structure including handle, head set, water and air connector. Inside the head set with turbine shaft,turbine,shaft. Inside the handle with air pipe,water pipe. The outlet of the turbine shaft is provided with a bur jack. The diameter of the bur is Φ1.595-Φ1.600mm.
Accessory
Bur Wrench
Attention
●This product is for one-time use, do not reuse it. Please dispose it in accordance with relevant provisions after usage.
●This product is specialized for oral treatment, do not use it in other ways.
●This product is high speed handpiece, be careful when you use it.
●This product is only for doctors.
●Every time you use it, please check first . If find anything unnormal, stop using immediately, and contact your supplier.
●The damaged handpiece will cause very high noise, using it for a long time will influence your audition.
●To avoid being damaged badly, be careful not to drop it or hit it.
●Please use a bur of ISO-standard.Φ1.595-Φ1.600mm, any other bent,damaged or non-concentric burs will cause cartridge damaged badly.
●While using a bur, make sure that the bur's rotate speed and size fit the handpiece correctly.
●Always securely mout a bur in the chuck, even while it is not in use.
●Make sure that the rotation has stopped completely before replacing the cartridge or the bur.
●Maintenance statement: On-site requirement is not available.