EDC 501 J is a multi-range, high accuracy DC voltage source and calibrator for use in applications up to 10 Volts DC.
1.1.3 Accuracy:±(0.002% of programmed value + 0.0005% or range +3μV). The accuracy isbased on the "Limit of Error" (or "Worst Case" Method. All other specificationsnoted hereafter, which effect accuracy, e.g., line, load, temperature, and drift areincluded in the accuracy statement. Thus, all other specifications listed are (*)NON-ADDITIVE.1.1.4 Settling Time:100 μs (50 μs nominal) to rated accuracy for voltage change. 10 ms (approx.)recovery from short-circuit.100 μs (max) for polarity change.60 ms (max) for range change.1.1.5 Stability (non-additive specifications)24 hrs: 0.001%; 90 days: ±0.0015%; 6 months: ±0.002%1.1.6 Ripple and Noise:Observe in a band pass of 0.1 Hz to 100 kHz with NO random spikes. (0.0005%of range +2 μV) rms.1.1.7 Output Current:10 V range: 50 mA100 mV range: Not Applicable. (Output voltage is within stated accuracy if loadimpedance is greater than 60 kΩ.)