Function:

1. It allows you to measure DC V, DC mV, resistance, TC and RTD.

2. Outputs resistance, simulation of TC and RTD.

3. Thermocouple (TC) measurement provides cold-junction compensation.

4. Manual stepping, automatic stepping, 0-100% phase step and ramp output.

5. Super-strong protection:The waterproof grade is IP67. Automatic protection against the signal terminals connect to the 220V.

6. Support for PC communication

 

Summary of Source and Measure Functions:

Function

Measure

Source

DC V

0-30V

Not available

DC mV

0-120mV

0mV-100mV

Resistance

0-3200Ω

0-3200Ω

RTD

Pt100, Pt1000, Cu50, Cu100

Pt100, Pt1000, Cu50, Cu100

TC

E,J,K,T,B,R,S,N

E,J,K,T,B,R,S,N

Others

Stepping output, ramp, phase step output, user-defined range

 

Specifications:

Specifications are based on a one year calibration cycle and apply from +18C to +28 C unless stated otherwise. 

All specifications assume a 10 minute warmup period.

DC Voltage Measurement

Range

Maximum measurement range

Resolution

Accuracy (% of reading + Counts)

ETX-1810

ETX-2010

30V

0V-31V

0.001V

0.05+2

0.02+2

100mV

-15mV-80mV

0.001mV

0.05+20

0.02+20

80mV-125mV

0.01mV

0.05+2

0.02+2

-10C-18C, +28C-55C temperature coefficient, ±0.005%FS/C

Input resistance: >1MΩ.

DC Voltage Source

Range

Maximum output range

Resolution

Accuracy (% of reading + Counts)

ETX-1810

ETX-2010

100mV

-15mV-99.999mV

0.001mV

0.05+20

0.02+20

100mV-125mV

0.01mV

0.05+2

0.02+2

-10C-18C, +28C-55C temperature coefficient, ±0.005%FS/C

Maximum load: 1mA or 1kΩ (It should be based on the lower load.)

Resistance Measurement

Range

Maximum measurement range

Resolution

Accuracy (Ω)

ETX-1810

ETX-2010

2-wire, 3-wire

4-wire

2-wire,

3-wire

4-wire

400Ω

0-440Ω

0.01Ω

0.25

0.15

0.15

0.10

3200Ω

420Ω-3600Ω

0.1Ω

1.5

1.0

1.0

0.5

-10C-18C, +28C-55C temperature coefficient, ±0.005%FS/C

Exciting current during measurement:

400Ω: 1.0mA±10%;

3200Ω: 0.2mA±10%;

Two-wire: Conductor resistance is excluded from errors.

Three-wire: Matching test line should be used. The total resistance of conductor should not be larger than 25Ω.

Resistance Output

Range

Maximum output range

Resolution

External exciting current

Accuracy (Ω)

ETX-1810

ETX-2010

400Ω

0-440Ω

0.01Ω

0.4mA-4.0mA

0.25

0.15

3200Ω

400-3600Ω

0.1Ω

0.1mA-0.5mA

1.0

0.50

-10C-18C, +28C-55C temperature coefficient, ±0.005%FS/C

Temperature, TC

Graduation

Range

Resolution

Accuracy (Ω)

ETX-1810

ETX-2010

J

-200C-0C

0C-1200C

0.1C

1.5C

1.0C

1.0C

0.7C

K

-200C-0C

0C-1370C

0.1C

1.8C

1.2C

1.2C

0.8C

T

-200C-0C

0C-400C

0.1C

1.8C

1.2C

1.2C

0.8C

E

-200C-0C

0C-950C

0.1C

1.5C

1.0C

0.9C

0.7C

R

-20C-0C

0C-500C

500C-1750C

0.1C

4C

2.5C

2C

2.5C

1.8C

1.4C

S

-20C-0C

0C-500C

500C-1750C

0.1C

4C

2.5C

2C

2.5C

1.8C

1.5C

B

600C-800C

800C-1000C

1000C-1800C

0.1C

3.5C

2.5C

2C

2.2C

1.8C

1.4C

N

-200C-0C

0C-1300C

0.1C

2.0C

1.2C

1.5C

0.9C

Errors of cold-junction compensation are not included in the table.

Accuracy of cold-junction compensation: 1.5C

Temperature, RTD

Graduation

Range

Resolution

Accuracy (C)

ETX-1810

ETX-2010

2-wire

3-wire

4-wire

Output

2-wire

3-wire

4-wire

Output

Pt100

-200C-850C

0.1C

0.7

0.4

0.7

0.4

0.3

0.3

Pt1000

-200C-650C

0.4

0.3

0.3

0.3

0.15

0.15

Cu50

-50C-150C

1.2

0.8

0.8

0.8

0.5

0.5

Cu100

-50C-150C

0.7

0.4

0.4

0.4

0.25

0.25

As for exciting current during measurement, please refer to resistance measurement function.

As for allowable external exciting current during output, please refer to resistance output function.

2-wire: Does not include lead resistance.

3-wire: Assumes matched leads with a total resistance not exceeding 25Ω.

 

Accessories:

Hard spot test leads (one set)

Alligator clip (one set)

One 12V/1A power adaptor

Optional accessories:

6 R03-model rechargeable batteries

1 Metal Box

Communication line