Introduction
The CMS50N Pulse Oximeter adopts Photoelectric Oxyhemoglobin Inspection
Technology in accordance with Capacity Pulse Scanning & Recording
Technology, which can be used to measure human oxygen saturation and pulse rate
through finger. The device is suitable for being used in family, hospital,
oxygen bar, community healthcare and physical care in sports, etc. (It can be
used before or after exercise, but not recommended to use during
exercising).
Features
1) Can measure SpO2 and Pulse Rate accurately
2) SpO2 and Pulse Rate display,Pulse Rate waveform and bar graph
display
3) Battery voltage low indication
4)Auto standby
5) The display mode can be changed(Option)
6) Screen brightness adjustable
7) Shut off automatically when no signal
Main performance
●Display Mode:0.96" TFT display.
Screen resolution: 160*80.
●SpO2 measurement range: 0%~100% (resolution: 1 %).
Accuracy: 70%~100%: ±2%.
0%~69%: unspecified.
●PR measurement range: 30bpm~250bpm (resolution: 1 bpm).
Accuracy: ±2 bpm or ±2%, whichever is greater.
●Accuracy under low perfusion: low perfusion 0.4 %, SpO2 accuracy: ± 4 %;
PR accuracy: ± 2 bpm or ± 2 %, whichever is greater.
●Light interference: Under normal and ambient light conditions, the SpO2
deviation ≤ 1%.
●Working current: ≤30mA
●Power supply: two “AAA” batteries
●Safety classification: type BF applied part
Accessories
1)One hanging rope;(optional)
2)One User Manual.
Physical characteristic
Dimension: 60mm(L) × 34mm(W) × 31mm(H)