Description:
Introducing our versatile 8-inch patient monitor designed for adult, pediatric, and neonate monitoring. With abundant monitoring functions, this device ensures comprehensive care. Monitor ECG, RESP, SpO2, PR, NIBP, and TEMP effortlessly on the intuitive 8'' TFT color LCD touch screen. Stay up-to-date with PVI (Pleth Variability Index) and export data conveniently via USB. Connect wirelessly or by wire for Nurse Call and easily link to CMS with the USB WIFI module. Experience exceptional digital SpO2 performance alongside human voice alarms. Store 720 hours of trend data and 12 hours of waveforms, while choosing from 16 languages for a personalized experience. Elevate patient care with this reliable 6-parameter signs monitor.
Packing List:
Package Include: 1×Patient Monitor 1×NIBP Cuff 1×SpO2 Probe 1×5 EKG?Cable 1×Temperature Probe 1×NIBP extension cable 5×Electrode 1×User Manual
ECG
:5 Leads, I, II, III, AVR, AVL, AVF, V
Gain:0.25, ´0.5, ´1, ´2
Heart rate:
Adult:15 ~ 300 bpm
Neonatal/Pediatric:15 ~ 350 bpm
Accuracy:1%
Resolution:1 bpm
Sensitivity:> 200 μV P-P
Differential Input Impedance:> 5 M ohm
Electrode offset potential:300mV
Leakage Current:< 10 μA
PACE pulse detect
Range:±2~±700mV
Width:0.1~2ms
rise time:10~100µs
PACE pulse rejection
Range:±2~±700mV
Width:0.1~2ms
rise time:10~100µs
ST measurement
Range;-2.0 ~ +2.0 mV
Accuracy:-0.8mV~+0.8mV:
±0.02mV or ±10%, which is greater
Respiration
Method:Impedance between RA-LL or RA-LA
Differential Input Impedance:>2.5 M ohm
Respiration Impedance Range:0.3~3Ω
Base Impedance Range:200Ω-2000Ω
Bandwidth:0.3 ~ 2 Hz
Gain:×0.25,×0.500,×1,×2,×4
Respiration Rate
Measurement Range
Adult:0 ~ 120 BrPM
Neonatal / Pediatric: 0 ~ 150 BrPM
Resolution: 1 BrPM
Accuracy
0~6 BrPM: unspecified
7~150 BrPM: ±2 BrPM or ±2%, the greater
Inspiriting current:< 300 µA RMS max
Apnea Alarm:10 ~ 40 s
Temperature
Channel:1
Measure and Alarm Range:0 ~ 50 °C
Resolution:0.1°C
Accuracy:± 0.2℃
NIBP
Pulse Rate Range:40 ~ 240 bpm
Measure and Alarm Range
Adult Mode
SYS:40 ~ 280 mmHg
DIA:10 ~ 220 mmHg
MEAN:20 ~ 240 mmHg
Pediatric Mode
SYS:40 ~ 220 mmHg
DIA :0 ~ 160 mmHg
MEAN:20 ~ 170 mmHg
Neonatal Mode
SYS:35 ~ 135 mmHg
DIA:10 ~ 100 mmHg
MEAN:20 ~ 110 mmHg
Static pressure accuracy:3mmHg
Resolution:1mmHg
Accuracy
Maximum Mean error:5mmHg
Maximum Standard deviation:8mmHg
Over pressure Protection
Adult:300 mmHg
Pediatric:240 mmHg
Neonatal:150 mmHg
SpO2
Measurement Range:0 ~ 100 %
Resolution:1 %
Accuracy
2 % for 70% ~ 100%
0% ~ 69% unspecified
Pulse Rate
Measure and Alarm Range:20~250bpm
Resolution:1bpm
Accuracy:3bpm
Payment
We accept PayPal only. We will ship out the item after receiving cleared payment.
Shipping
A complete and correct shipping address is essential.
Customers with shipping addresses outside of the U.S. are responsible
for all duties and import taxes. Please contact your customs for
information
Return Policy
All items are in brand new condition unless specified.
FDA declaration:
The sale of this item may be subject to regulation by the U.S. Food and
Drug Administration and state and local regulatory agencies. If so, do
not bid on this item unless you are an authorized purchaser.If the item
is subject to FDA regulation, we will verify your status as an
authorized purchaser of this item before shipping of the item.All of our
products are CE approved. Please kindly check the copy of the
certificate in the listing below.(Rose-China-Beijing-86-13578947868)
This item has been cleaned and treated according to the manufacturer's instructions.
510(K)Number: K161892
Contact Us
Any questions, please contact us through eBay message or e-mail, We
will reply within 1-2 business days (public holidays not included).
working time: Monday - Friday 9:00 AM - 5:30PM (Beijing)