Introduction
ECG600G is an Electrocardiograph which can collect 12-lead ECG signal
simultaneously and print waveform by the thermal printing system. It features
in, recording and displaying ECG waveform in AUTO/Manual mode, measuring and
diagnosing ECG waveform parameters automatically, prompting for “Lead off” and
“Lack of paper”, multi-language interface, case database management.
Function
1)Adopt
high-resolution thermal printing system(8 dot/mm), no need for any adjustments.
Recording frequency: up to 150Hz.
2)Real-time and continuously record clear
and accurate ECG waveform and annotation character(including lead mark,
sensitivity, paper speed and filter state, etc.).
3)Full-digital filter
design, anti-baseline drift, AC and EMG interference.
4)Adjust the baseline
automatically, optimize the print setting and optional rhythm leads.
5)In
AUTO mode, finish recording by pressing the button once, which improves work
efficiency.
6)Adopt digital signal processing technology and get high-quality
ECG waveform via power frequency filter, baseline filter and EMG filter of ECG
signal.
7)With the functions of auto-measurement, auto-analysis and
auto-diagnosis for routine ECG parameters, which reduces the doctor’s burden and
improves work efficiency.
8)Sync collection for 12-lead ECG, multiply
printing formats, including 2×6+1(rhythm lead), 2×6, 3×4, 3×4+1(rhythm lead),
4×3, 4×3+1(rhythm lead), 6×2, 6×2+1(rhythm lead), rhythm 2, rhythm 3, rhythm 4,
rhythm 5, rhythm 6, rhythm 7, channel record, etc.
9)Display of real-time
3/6/12-lead ECG, "Lead off" and "Lack of paper" on one screen, and with ECG DEMO
mode. Auto-check and auto-analysis for ECG parameters, ECG waveform can be
recorded manually or automatically. And patient information can be stored by
case database.
10)Time print can be set to every 1/2/3/5/10/20/30/60
minutes.
11)Print setting, including “Print and save”(print waveform and save
case), “Print no save”(only print waveform but not save case), “Save no
print”(only save case but not print waveform).
12)AC/DC, built-in
rechargeable lithium battery and auto-protection circuit, and automatic power
off when there is no operation in appointed time.
13)Store more than 1000
cases, convenient for case review and statistic.
14)Multi-language(Chinese,
English, French, Spanish, Turkish, Polish, Italian, German, Portuguese, Kazakh,
Russian, Ukrainian, Vietnamese, Serbian, Slovenian and Slovak) interface and
report.
Performance
Input mode:
floating and defibrillation protection
Input impedance: ≥2.5 MΩ
Patient
leakage current: <10 µA
Frequency response:
Rated input amplitude
Input frequency and waveform Relative output response
1.0 0.67Hz~40Hz,
Sine wave ±10%a
0.5 40Hz~100Hz, Sine wave +10 %, -30 %a
0.25
100Hz~150Hz, Sine wave +10 %, -30 %a
0.5 150 Hz ~ 500 Hz, Sine wave
+10 %, -100 %a
1.5 ≤1Hz,200ms, Triangle wave +0 %, -10 %b
a
relative to 10Hz b relative to 200 ms
Time constant:≥ 3.2s
CMRR:
>105dB
Filter: power frequency(AC50/60 Hz), EMG(25 Hz/35 Hz (-3 dB)),
baseline drift filter
Recording way: thermal printing system
Specification
of recording paper: 110 mm(W)×20 m(L) high-speed thermal paper
Time base
selection(paper speed): 12.5 mm/s, 25 mm/s, 50 mm/s, error: ±5%
Gain
control(sensitivity): 5, 10, 20 mm/mV, accuracy is ±2%; Standard sensitivity: 10
mm/mV±0.2 mm/mV
Recording mode: auto mode, rhythm mode and manual
mode
Measurement parameters: HR, P-R interval, P Duration, QRS Duration, T
Duration, Q-T interval, Q-Tc, P Axis, QRS Axis, T Axis, R(V5), S(V1),
R(V5)+S(V1)
Safety classification: Class I, type CF defibrillation-proof
applied part
Polarization resistance voltage: ±610mV
Noise level: ≤12
µVp-p
ECG signal input sampling frequency: 32 kHz
Waveform data processing
sampling frequency: 1 kHz
Sampling accuracy: 24-bit
The minimum detection
signal: 10 Hz, 20 µV(peak-peak value) deflected sinusoidal signal can be
detected
Pacing detection channel: standard II
Accuracy of input signal:
±5%
Amplitude quantization: ≤5µV/LSB
Interchannel time deviation: <100
µs
Power supply:
AC: 100V~240V(50/60Hz)
DC: 7.4V/3500mAh rechargeable
lithium battery
Degree of protection against ingress of liquid:
IPX0
Working mode: continuous working
Accessories
Standard
configuration:
A lead cable
A limb electrode
A chest electrode
A
thermal recording paper
A power cord
An earth wire
A user manual
Physical
characteristic
Working environment
Temperature:
5℃~40℃
Relative humidity: 25%~95%(non-condensing)
Atmospheric pressure:
700 hPa~1060 hPa
Transport and storage environment
Temperature: -20 ℃~+55
℃
Relative humidity: ≤95%
Atmospheric pressure: 500 hPa~1060
hPa
Dimension: 315mm(L)×215mm(W)×92mm(H)
Weight: 1.7Kg
Certification
SFDA, CMC,
CE.