TS13  Patient Monitor

           
               
                   
Introduction
The monitor is applicable for clinical monitoring in operating room, postoperative observation room, ICU/CCU ward, emergency room, etc. of adult, pediatric and neonate. The patient's vital parameters (ECG (including ST-segment measurement and arrhythmia analysis), RESP, SpO2, PR, NIBP, TEMP, IBP and CO2, etc.) can be monitored. Monitored information can be displayed, reviewed, printed and stored..

Function
1)13.1" high-definition touch screen design, easy and convenient to operate, elegant and concise in appearance.
2)Independent physiological and technical alarm function, convenient for medical staff to quickly know patient’s condition.
3)Luminous “ON / OFF” button design, convenient to operate at night.
4)Modular design: the modules can be configured and combined flexibly and conveniently, expanded and upgraded at any time, meets the monitoring requirements of different departments on different parameters.
5)Support dynamic hot-plugging parameter module: no need to restart the device, plug-and-play, change the monitoring parameters at any time as clinical requirements.
6)Adopt high-performance processor, stable, reliable and fast to process information.
7)Patient information can be rapidly entered, patient type (adult / neonate / pediatric) can be quickly switched, meets the requirements of different departments.
8)Abundant analysis functions: 23 arrhythmia, full-lead ST-segment and pacing analysis; multiple calculation functions (such as calculation of drug concentration, titration table, hemodynamics, ventilation, oxygenation, renal function, etc.).
9)With the functions of waveform freezing, holographic waveform storage and review.
10)Automatic data storage in case of power failure; optional built-in memory card, achieves large-capacity data storage.
11)With HDMI expansion interface, the display screen can be connected according to clinical requirements.
12)Built-in large-capacity detachable lithium battery, ensures uninterrupted monitoring.
13)Fanless design, ensures no noise and low power consumption during working.
14)3-channel built-in recorder used to print real-time waveform is optional.
15)Connect to Central Monitoring System developed by our company by WiFi / Wired, convenient to transmit patient’s real-time data.
16)Standby mode, be applied to ICU, avoids affecting patient’s rest, one-button switching, reduces the workload of medical staff.

Performance
(1)ECG
  Lead type:               
3-lead: I, II, III
5-lead: I, II, III, aVR, aVL, aVF, V
12-lead: I, II, III, aVR, aVL, aVF, V1,V2,V3,V4,V5,V6
 Gain: 2.5 mm/mV, 5.0 mm/mV, 10 mm/mV, 20 mm/mV, 40 mm/mV
  Scan speed: 12.5 mm/s, 25 mm/s, 50 mm/s
  HR
Measurement and alarm range:   
Adult: 15 ~ 300 bpm
Pediatric/neonate: 15 ~ 350 bpm
Resolution: 1 bpm
Accuracy: ± 1 % or ± 1 bpm, whichever is greater
  ST-segment detection
Measurement and alarm range: - 2.0 mV ~ + 2.0 mV
Accuracy: - 0.8 mV ~ + 0.8 mV: ± 0.04 mV or ± 10 %, whichever is greater
Other range: unspecified
  Arrhythmia analysis: 23 kinds
  Pacemaker: yes
(2)RESP
  Measurement mode: R-F (RA-LL) impedance
  Respiration rate:
Measurement range:
Adult: 0 rpm ~ 120 rpm
Pediatric/neonate: 0 rpm ~ 150 rpm
Resolution: 1 rpm
Accuracy: 7 ~ 150 rpm, ± 2 rpm or ± 2 %, whichever is greater;
 0 ~ 6 rpm, unspecified
  Apnea alarm: 10 ~ 120 s, not alarm
  Scan speed: 6.25 mm/s, 12.5 mm/s, 25 mm/s, 50mm/s
(3)NIBP
  Measurement mode: Oscillometry
  Working mode: Manual/AUTO/Continuous
  Measurement interval in AUTO mode: 1 min, 2 min, 2.5 min, 3 min, 5 min, 10 min, 15 min, 20 min, 30 min, 1 h, 1.5 h, 2 h, 3 h, 4 h, 8 h
  Measurement period in Continuous mode: 5 minutes
  Measurement and alarm range: 10 ~ 270 mmHg
  Resolution: 1 mmHg
  Cuff pressure accuracy: ± 3 mmHg
  Measurement accuracy:
Maximal mean deviation: ± 5 mmHg
Maximal standard deviation: 8 mmHg
  Over-pressure protection:
Adult mode: 297 ± 3 mmHg
Pediatric mode: 240 ± 3 mmHg
Neonatal mode: 147 ± 3 mmHg
(4)CONTEC SpO2
  Measurement range: 0 % ~ 100 %
  Resolution: 1 %
  Accuracy: 70 % ~ 100 %, ± 2 %;
 0 % ~ 69 %, unspecified
  PR
Measurement and alarm range: 25 ~ 250 bpm
Accuracy: ± 2 bpm or ± 2 %, whichever is greater
(5)Nellcor SpO2
  Measurement range: 1 % ~ 100 %
  Measurement accuracy: 70 % ~ 100 %: ± 2 % (adult / pediatric)
 70 % ~ 100 %: ± 3 % (neonate)
 0 % ~ 69 %: unspecified
  Nellcor PR
Measurement range: 20 ~ 300 bpm;
Measurement accuracy: 20 ~ 250 bpm, ±3 bpm;
 251 ~ 300 bpm: unspecified;
Upper limit of alarm: upper limit - 2 ~ 300 bpm;
Lower limit of alarm: 20 bpm ~ lower limit + 2 bpm
(6)Masimo SpO2
  Measurement range: 1 % ~ 100 %
  Measurement accuracy:
70 % ~ 100 %: ± 2 % (adult / pediatric, under conditions of no motion), ± 3 % (under conditions of motion)
70 % ~ 100 %: ± 3 % (neonate, under conditions of no motion)
1 % ~ 69 %: unspecified
  Masimo PR
Measurement range: 25 ~ 240 bpm
Measurement accuracy: ± 3 bpm (under conditions of no motion), ± 5 bpm (under conditions of motion)
Upper limit of alarm: upper limit - 2 ~ 240 bpm
Lower limit of alarm: 25bpm ~ lower limit + 2 bpm
(7)TEMP
  Channel: dual-channel
  Measurement range: 0 ~ 50 ℃
  Resolution: 0.1 ℃
  Accuracy: ± 0.1 ℃
(8)CO2
  Measurement mode: mainstream or sidestream
  CO2 measurement range: 0 ~ 150 mmHg
  CO2 resolution: 0.1 mmHg: 0 ~ 69 mmHg
 0.25 mmHg: 70 ~ 150 mmHg
  Accuracy:
0 ~ 40 mmHg: ± 2 mmHg
41 ~ 70 mmHg: ± 5 %
71 ~ 100 mmHg: ± 8 %
101 ~ 150 mmHg: ± 10 %
  AwRR measurement range:
Sidestream: 2 ~ 120 rpm
Mainstream: 2 ~ 150 rpm
  AwRR accuracy: ± 1 rpm
  Apnea alarm: yes
(9)IBP
  Channel: 4-channel
  Label: ART, PA, CVP, RAP, LAP, ICP, P1, P2
  Measurement and alarm range: -50 ~ 300 mmHg
  Resolution: 1 mmHg
  Accuracy: ± 2 % or 1 mmHg, whichever is greater
(10)Power supply: 100 ~ 240 V ~  50/60 Hz
(11)Safety class: class I, defibrillation-proof type CF applied part
(12)Waterproof degree: IPX2

About Us

Contec Medical Systems focusing on research, manufacture and distribution of medical instruments,was founded in 1992 as a high-tech company.At present there are more than 1200 employees in our company.Our product line covers a wide range of 13 categories.Most of the domestic hospitals are our customers.Contec hopes to cooperate with international companies to supply more innovative design and advanced technology products We sincerely welcome you to become one of our global partners.We are looking forward to establishing a successful business relationship with you.

Shipping
1.Our shipping rates include the cost of shipping, handling, packaging,custom fees are not included,buyers pay the taxes.
2.We usually shipped out the items the next business day after payment has been verified (Sat. & Sun. excluded.)

Contact Us
Please contact us first if you are not satisfied with the item when you receive it or haven't received item after 30 days, We will try our best to solve any problem. Open case and negative feedback is not conducive to solve the problem for both sides. hope you can understand!
All emails will be answered within 24 hours on weekdays(but within 48hours on weekends).Please don't hesitate to contact us with any queries or comments about our items or your problem.
Only English operating instruction,if you need any other language,please contact us!
                    
               
                                
               


   
































































               

           
       
       
   