ORP redox potentiometer ORP electrode measures redox potential new Description: ORP (also known as redox potential) composite electrode is a composite electrode composed of a gold or platinum electrode and a reference electrode, which can measure the redox potential of a solution. The main technical parameters: Product Name Model Plastic ORP Composite Electrode-Type 501 Feature PC case, no need to refill gel fluoride Reference electrode (built-in) Ag / Agcl Liquid junction ceramic core Dimensions φ12 * 70mm Wire length 1m BNC plug MV value in ORP standard solution: 256 ± 15mV (25 ℃) Color: black blue green (random) The test in ORP standard fluid 1. Connect the ORP electrode to the pH / mV meter, and set the electronic unit to the mV measurement range. 2 immerse the ORP composite electrode in the 256MV standard solution, leave it agitated, and leave it to rest. After the displayed value stabilizes, the reading should meet the above technical standards. Package includes: 1x Redox Electrode
|