Tensiometer Test Pump Service Unit 1007

Service Units are designed to aid users in the placement and installation, interpretation of readings, and maintenance of IRROMETER soil moisture indicators. Several helpful items are included in the kit. The primary component of a Service Unit is a small hand vacuum pump. The pump is designed to seal on the reservoir portion of the instrument body and is used to create a vacuum manually, evacuating any air contained in the tensiometer. These instruments function with the greatest accuracy when all air has been removed. One Service Unit will service multiple instruments.  This is used to validate accuracy of the gauge on the IRROMETER being serviced.

Pump Features: • Designed for outdoor field use • Non-corrosive construction • Easily serviceable • Long lasting

OPERATING PRINCIPLE: The Service Units provide the IRROMETER user with valuable tools to ensure their soil moisture indicators function accurately and are easily operated and maintained. The hand vacuum pump serves to evacuate entrapped air from the IRROMETER body which optimizes accuracy and response time. This is important for initial installation and also for periodic maintenance purposes, as air may be drawn into the instrument from the soil under normal operating conditions. Viewing of the fluid level in the instrument and refilling is made easier by using the IRROMETER Fluid concentrate to color the fill water. Included Chart Forms provide a convenient way to graph the periodic IRROMETER readings for use when analyzing soil moisture trends. The instruction booklet provides detailed information on unpacking, preparing, installing, servicing and storing IRROMETERS as well as guidance on instrument placement and data interpretation.

MODEL “MLT” FILL & AIR REMOVAL INSTRUCTIONS — 1. Remove cap, tip protector and plastic bag covering tip. 2. Place tip in clean water and soak overnight – you do NOT need to fill the instrument. Step 5 below will complete the filling process. 3. Make IRROMETER field solution by adding one drop of IRROMETER fluid to a cup of clean water. 4. Fill syringe half full with IRROMETER field solution. 5. To fill, place syringe tip loosely in O-ring seal (located in cap opening), depress syringe plunger to squirt water into the body tube of the MLT. Do not seat the syringe to the Oring in this step as it could force the gauge needle against the stop. (RSU units do NOT have a gauge needle). 6. To remove air, place syringe tip in O-ring seal (located in cap opening), press in firmly to make a seal. With instrument at 45 angle (gauge or RSU down or on the underside), PULL GENTLY on syringe to create a partial vacuum on the gauge or RSU (on gauge model, do NOT exceed full scale of gauge or 40 kPa/CB) and release slowly. Pumping this way several times will get most of the air out of the gauge or RSU. 7. Remove syringe by slowly twisting while pulling out to avoid a rapid release of vacuum, fill instrument cap opening with water. 8. Replace cap – tighten until gauge needle moves toward De-airing in the Ground De-airing Instrument De-airing MLT Filling MLT 4 stop. 9. Tip (body) of instrument can be removed by pulling out from the top housing. Be sure to remove the cap before attempting to pull the tip (body) out to avoid damaging the gauge. Replacement tips (bodies) are available.