Funtion
This 2-in-1 WIFI pH & ORP controller and monitor simultaneously measures and displays both pH value and redox value (ORP). It comes with 2 BNC electrodes (pH and redox) and 2 output power sockets. Both electrodes are detachable, they can be replaced easily at any time. As for the relay output power sockets, you can plug in your own 110V/ 220V device to it, e.g. air outlet stone, O3 ozoniser, oxygen pump, CO2-regulator, electromagnetic valve and other ORP & pH control devices. Simply set the desired Hi/ Low pH/redox values in the controller. It will turn on your device if the currently measured pH or Redox value is lower than your set value, and it will turn off your device if the measured value is higher than your set value.
Operation and Clibration:
Calibration steps:
pH calibration:
In pH6.86 Calibration solution
1.Connecting AC power supply. Immerse the ph electrode in a pH6.86 buffer solution, and gently shake it until the reading to stabilize.
2.Long press [b] ,until the indicator light of “pH” flashes.PH6.86 is shown in the first line.
3.(1)Short press [a] to exit the current mode if the calibration of ph6.86 is not performed.Then the device returns to measurement mode.
(2)Short press [c] to complete the calibration of pH 6.86 when the reading of the second line is stable if the calibration of ph6.86 is performed.Then the device automatically returns to measurement mode.
4.The calibration of pH6.86 is completed.
In pH4.00 Calibration solution
1.Connecting AC power supply.Immerse the ph electrode in a pH 4.00 buffer solution, and gently shake it until the reading to stabilize.
2.Long press [e],until the indicator light of “pH” flashes.pH 4.00 is shown in the first line.
3.(1)Short press [a] to exit the current mode if the calibration of pH 4.00 is not performed.Then the device returns to measurement mode.
(2)Short press [c] to complete the calibration of pH 4.00 when the reading of the second line is stable if the calibration of ph4.00 is performed.Then the device automatically returns to measurement mode.
4.The calibration of pH 4.00 is completed.
ORP calibration:
This meter has been accurately precalibrated for mV range at the factory. and there is no need for users to calibrate.
Connecting device:
Ensure the device and phone connect the same WIFI.
Register and log in.
In the interface of My Device,click “…” in the upper right corner,and click Add Device.
Enter the WiFi ID and the password.Then click Next.
Enter the interface of module type selection.Click ESP.
Long press [a] of the ph-803W until the WiFi signal flashes.Check "Steps have been completed" and click Next.
After reaching 100%,you can check the device successfully connected to the Internet in the interface of My devices.Clicking the device,you can view the reading of pH and ORP as well as the on-off condition of the power control box.
After the device is connected to the Internet, click "... "in the upper right corner. , you can rename the device in the interface of Settings Device Information.You can check the readings of your bound device anywhere through the phone.
Attention:
Calibrating again after long time break
Keep the water not over the probe.
Application:
This device is suitable for aquarium,hydroponics planting,fish farming and tank and pool.
Descriptio:
Measuring range
PH: 0.00 ~ 14.00pH
ORP: -1999 ~ 1999mV
Accuracy
PH: plusmn0.1%+0.02PH
ORP: plusmn5mV
Resolution
PH: 0.01pH
ORP: 1mV
Set Range
PH: 3.50 ~ 10.50pH
ORP: -500 ~ 500mV
Relay Contact: 5A for AC 110V or 220V
Environment
Operation Temperature : 0 to 50 degC (32 to 122 degF)
Ambient Humidity : 90 or less than 90 % R. H.
Power Requirement : AC110V plusmn10% 60Hz / AC220V plusmn10% 50Hz
Power Consumption : 2 Watts
Voltage: 110V or 220V
Precautions: