Lab Equipment Smart Disintegration tester BJ-3 110V/220V
BJ series of instruments are used for detecting disintegration of solid in prescriptive conditions
Applicable standard
* National standard (Chinese codex)
* Corporate standard (Disintegration tester) Q/12xQ0188-2004
Parameters:
Number of lift nacelle: 3 nacelle tube number: 18 flat beaker volume: 1000ml
Nacelle back and forth frequency: 30~32 times per minute
Nacelle back and forth range: 55±1mm
Minimum distance between net and glass's bottom: 25±2mm Time limit warning range: 1minute~9hours59minutes(set at will)
timing accuracy: ± 0.5min temperature preset range: (20 ~ 45) ℃
automatic temperature control accuracy: ± 0.5 ℃ continuous working time: more than 24 hours
Baffle accord with Chinese codex in 2000
Rated voltage: 220V 110V Rated power: 600W
Size: 540*320*440mm
Weight: 22kg |
The Frequency can be 50HZ or 60HZ. Different frequency will affect the speed.
If you have no special requirement, we will make 220V 50HZ or 110V 60HZ by default.
If you need 220V 60HZ or 110V 50HZ, please contact us.