TECHNICAL SPECIFICATIONS:
Mode |
General |
Serum |
Urine |
Range |
Na:
0 – 100 ppm Upto
250 meq/L 1: 100 dil |
Na:
0 – 200 1:100
dil |
Na:
0-250 1:100
dil |
|
K:
0-100 ppm Upto
250 meq/L, 1: 100 dil |
K:
0-10 1:100
dil |
K:
0-100 1:100
dil |
|
Ca:
0-100 ppm Upto
250 meq/L, 1: 100 dil |
- |
Ca:
0-10 1:2
dil |
|
Li:
0-100 ppm Upto
250 meq/L, 1:100 dil |
Li:0-2 1:5
dil |
- |
Sensitivity |
General:
Na : 0.5 ppm, K : 0.5ppm, Li: 0.5 ppm, Ca: 15 ppm |
||
Filters |
Two Filters – Sodium (Na) & Potassium
(K) |
||
Resolution |
0.1
ppm/meq |
||
Reproducibility |
<
1% coefficient of variation for 20 consecutive samples using 10 ppm Na set as
maximum standard |
||
Curve
Fit Accuracy |
<
2% error when 3 ppm Na/K and 5 ppm Li are set as max std. |
||
Display |
128
x 64 Graphical LCD |
||
Average
Time |
In
built in software |
||
Flame
System |
LPG
and dry oil free air |
||
Detector |
Silicon
Photodiode |
||
Calibration |
Up
to 5- point calibration with curve fitting software |
||
Gas
Control |
Adjustable
with knobs |
||
Flame
Failure |
Auto
detection |
||
Gas
Cutoff |
Automatic
on power failure |
||
Ignition
System |
Auto
ignition |
||
Printer |
Provision
for attachment of dot matrix printer |
||
RS232C |
Provision
for data transfer to PC |
||
Power |
230
V +10% AC, 50 Hz |
||
Dimensions |
285
x 255 x 210 mm (L x B x H) approx. |
||
Weight |
7.5
kg approx. |
||
COMPRESSOR UNIT |
|||
Air
Supply |
By
oil free mini compressor unit with pressure regulator |
||
Combustion |
LPG
controlled by precision regulator |
||
Power |
230
V +10% AC, 50 Hz |
||
Dimensions |
290
x 255 x 210 mm (L x B x H) approx. |
||
Weight |
8
Kg approx. |
||
STANDARD ACCESSORIES |
In-built Filters (2), Sample Holders-5 Nos.,
Air & Gas Tubing-1 Set, Capillary Tubes-3 Nos., Mains Lead-1 No., RS-232
Cable-1 No. (Optional Wherever Printer Interface is provided) & Operation
Manual. |
Appearance
& Specifications are subject to change due to continuous improvements.